"product_name":"GE CARESCAPE Telemetry Server,ApexPro Telemetry Server,CARESCAPE Central Station,Clinical Information Center systems,CARESCAPE B450,B650,B850 Monitors",
"version":{
"version_data":[
{
"version_value":"ApexPro Telemetry Server,v4.2 & prior,CARESCAPE Telemetry Server, v4.2 & prior,Clinical Information Center,v4.X& 5.X,CARESCAPE Telemetry Server,v4.3,CARESCAPE Central Station,v1.X,CARESCAPE Central Station,v2.X,B450,v2.X,B650,v1.X,B650,v2.X,B850,v1.X,B850,v2.X"
}
]
}
}
]
}
}
]
}
},
"problemtype":{
"problemtype_data":[
{
"description":[
{
"lang":"eng",
"value":"UNRESTRICTED UPLOAD OF FILE WITH DANGEROUS TYPE CWE-434"
"value":"In ApexPro Telemetry Server Versions 4.2 and prior, CARESCAPE Telemetry Server v4.2 & prior, Clinical Information Center (CIC) Versions 4.X and 5.X, CARESCAPE Central Station (CSCS) Versions 1.X, B450 Version 2.X, B650 Version 1.X, B650 Version 2.X, B850 Version 1.X, B850 Version 2.X, a vulnerability in the software update mechanism allows an authenticated attacker to upload arbitrary files on the system through a crafted update package."