"CVE_data_meta": {
"ID": "CVE-2015-1012",
2019-03-25 19:00:51 +00:00
"ASSIGNER": "ics-cert@hq.dhs.gov",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"vendor_name": "Hospira",
"product": {
"product_data": [
{
"product_name": "LifeCare PCA Infusion System",
"version": {
"version_data": [
{
"version_value": "<= 5.0"
}
]
}
}
]
}
}
]
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Cleartext storage of sensitive information CWE-312"
}
]
}
]
},
"references": {
"reference_data": [
{
"refsource": "MISC",
"name": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01",
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-125-01"
}
]
2019-03-18 03:59:39 +00:00
},
"description": {
"description_data": [
{
"lang": "eng",
2019-03-25 19:00:51 +00:00
"value": "Wireless keys are stored in plain text on version 5 of the Hospira LifeCare PCA Infusion System. According to Hospira, version 3 of the LifeCare PCA Infusion System is not indicated for wireless use, is not shipped with wireless capabilities, and should not be modified to be used in a wireless capacity in a clinical setting. Hospira has developed a new version of the PCS Infusion System, version 7.0 that addresses the identified vulnerabilities. Version 7.0 has Port 20/FTP and Port 23/TELNET closed by default to prevent unauthorized access."
2019-03-18 03:59:39 +00:00
}
]
}
}