cvelist/2017/12xxx/CVE-2017-12721.json

68 lines
2.0 KiB
JSON
Raw Normal View History

2017-10-16 12:31:07 -04:00
{
"CVE_data_meta" : {
2018-02-15 05:04:12 -05:00
"ASSIGNER" : "ics-cert@hq.dhs.gov",
2017-10-16 12:31:07 -04:00
"ID" : "CVE-2017-12721",
2018-02-15 05:04:12 -05:00
"STATE" : "PUBLIC"
},
"affects" : {
"vendor" : {
"vendor_data" : [
{
"product" : {
"product_data" : [
{
"product_name" : "Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump",
"version" : {
"version_data" : [
{
"version_value" : "Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump"
}
]
}
}
]
},
"vendor_name" : "n/a"
}
]
}
2017-10-16 12:31:07 -04:00
},
"data_format" : "MITRE",
"data_type" : "CVE",
"data_version" : "4.0",
"description" : {
"description_data" : [
{
"lang" : "eng",
2018-02-15 05:04:12 -05:00
"value" : "An Improper Certificate Validation issue was discovered in Smiths Medical Medfusion 4000 Wireless Syringe Infusion Pump, Version 1.1, 1.5, and 1.6. The pump does not validate host certificates, leaving the pump vulnerable to a man-in-the-middle (MITM) attack."
}
]
},
"problemtype" : {
"problemtype_data" : [
{
"description" : [
{
"lang" : "eng",
"value" : "Improper Certificate Validation"
}
]
}
]
},
"references" : {
"reference_data" : [
{
"name" : "https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A",
"refsource" : "MISC",
2018-02-15 05:04:12 -05:00
"url" : "https://ics-cert.us-cert.gov/advisories/ICSMA-17-250-02A"
2018-02-15 06:04:25 -05:00
},
{
"name" : "100665",
"refsource" : "BID",
2018-02-15 06:04:25 -05:00
"url" : "http://www.securityfocus.com/bid/100665"
2017-10-16 12:31:07 -04:00
}
]
}
}