"data_type": "CVE",
|
||||
"CVE_data_meta": {
|
||||
"ASSIGNER": "productcert@siemens.com",
|
||||
"ID": "CVE-2020-15797",
|
||||
"STATE": "PUBLIC"
|
||||
},
|
||||
"data_format": "MITRE",
|
||||
"data_version": "4.0",
|
||||
"CVE_data_meta": {
|
||||
"ID": "CVE-2020-15797",
|
||||
"ASSIGNER": "cve@mitre.org",
|
||||
"STATE": "RESERVED"
|
||||
"data_type": "CVE",
|
||||
"affects": {
|
||||
"vendor": {
|
||||
"vendor_data": [
|
||||
{
|
||||
"vendor_name": "Siemens",
|
||||
"product": {
|
||||
"product_data": [
|
||||
{
|
||||
"product_name": "DCA Vantage Analyzer",
|
||||
"version": {
|
||||
"version_data": [
|
||||
{
|
||||
"version_value": "All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797"
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
},
|
||||
"problemtype": {
|
||||
"problemtype_data": [
|
||||
{
|
||||
"description": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "CWE-269: Improper Privilege Management"
|
||||
}
|
||||
]
|
||||
}
|
||||
]
|
||||
},
|
||||
"description": {
|
||||
"description_data": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
|
||||
"value": "A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Improper Access Control could allow an unauthenticated attacker to escape from the restricted environment (\u201ckiosk mode\u201d)\nand access the underlying operating system. Successful exploitation requires direct physical access to the system. \n"
|
||||
}
|
||||
]
|
||||
},
|
||||
"references": {
|
||||
"reference_data": [
|
||||
{
|
||||
"refsource": "CONFIRM",
|
||||
"url": "https://www.siemens-healthineers.com/support-documentation/security-advisory"
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
}
|
||||
|
@ -1,18 +1,61 @@
|
||||
{
|
||||
"data_type": "CVE",
|
||||
"CVE_data_meta": {
|
||||
"ASSIGNER": "productcert@siemens.com",
|
||||
"ID": "CVE-2020-7590",
|
||||
"STATE": "PUBLIC"
|
||||
},
|
||||
"data_format": "MITRE",
|
||||
"data_version": "4.0",
|
||||
"CVE_data_meta": {
|
||||
"ID": "CVE-2020-7590",
|
||||
"ASSIGNER": "cve@mitre.org",
|
||||
"STATE": "RESERVED"
|
||||
"data_type": "CVE",
|
||||
"affects": {
|
||||
"vendor": {
|
||||
"vendor_data": [
|
||||
{
|
||||
"vendor_name": "Siemens",
|
||||
"product": {
|
||||
"product_data": [
|
||||
{
|
||||
"product_name": "DCA Vantage Analyzer",
|
||||
"version": {
|
||||
"version_data": [
|
||||
{
|
||||
"version_value": "All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797"
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
},
|
||||
"problemtype": {
|
||||
"problemtype_data": [
|
||||
{
|
||||
"description": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "CWE-259: Use of Hard-coded Password"
|
||||
}
|
||||
]
|
||||
}
|
||||
]
|
||||
},
|
||||
"description": {
|
||||
"description_data": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
|
||||
"value": "A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an\nattacker to read and or modify the onboard database. Successful exploitation requires direct physical\naccess to the device.\n"
|
||||
}
|
||||
]
|
||||
},
|
||||
"references": {
|
||||
"reference_data": [
|
||||
{
|
||||
"refsource": "CONFIRM",
|
||||
"url": "https://www.siemens-healthineers.com/support-documentation/security-advisory"
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
}
|
||||
|