@ -4,14 +4,58 @@
|
||||
"data_version": "4.0",
|
||||
"CVE_data_meta": {
|
||||
"ID": "CVE-2020-27256",
|
||||
"ASSIGNER": "cve@mitre.org",
|
||||
"STATE": "RESERVED"
|
||||
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
||||
"STATE": "PUBLIC"
|
||||
},
|
||||
"affects": {
|
||||
"vendor": {
|
||||
"vendor_data": [
|
||||
{
|
||||
"vendor_name": "n/a",
|
||||
"product": {
|
||||
"product_data": [
|
||||
{
|
||||
"product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A",
|
||||
"version": {
|
||||
"version_data": [
|
||||
{
|
||||
"version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0"
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
},
|
||||
"problemtype": {
|
||||
"problemtype_data": [
|
||||
{
|
||||
"description": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "USE OF HARD CODED CREDENTIALS CWE-798"
|
||||
}
|
||||
]
|
||||
}
|
||||
]
|
||||
},
|
||||
"references": {
|
||||
"reference_data": [
|
||||
{
|
||||
"refsource": "MISC",
|
||||
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01",
|
||||
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01"
|
||||
}
|
||||
]
|
||||
},
|
||||
"description": {
|
||||
"description_data": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
|
||||
"value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, a hard-coded physician PIN in the physician menu of the insulin pump allows attackers with physical access to change insulin therapy settings."
|
||||
}
|
||||
]
|
||||
}
|
||||
|
@ -4,14 +4,58 @@
|
||||
"data_version": "4.0",
|
||||
"CVE_data_meta": {
|
||||
"ID": "CVE-2020-27258",
|
||||
"ASSIGNER": "cve@mitre.org",
|
||||
"STATE": "RESERVED"
|
||||
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
||||
"STATE": "PUBLIC"
|
||||
},
|
||||
"affects": {
|
||||
"vendor": {
|
||||
"vendor_data": [
|
||||
{
|
||||
"vendor_name": "n/a",
|
||||
"product": {
|
||||
"product_data": [
|
||||
{
|
||||
"product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A",
|
||||
"version": {
|
||||
"version_data": [
|
||||
{
|
||||
"version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0"
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
},
|
||||
"problemtype": {
|
||||
"problemtype_data": [
|
||||
{
|
||||
"description": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "INSUFFICIENTLY PROTECTED CREDENTIALS CWE-522"
|
||||
}
|
||||
]
|
||||
}
|
||||
]
|
||||
},
|
||||
"references": {
|
||||
"reference_data": [
|
||||
{
|
||||
"refsource": "MISC",
|
||||
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01",
|
||||
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01"
|
||||
}
|
||||
]
|
||||
},
|
||||
"description": {
|
||||
"description_data": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
|
||||
"value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, an information disclosure vulnerability in the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows unauthenticated attackers to extract the pump\u2019s keypad lock PIN via Bluetooth Low Energy."
|
||||
}
|
||||
]
|
||||
}
|
||||
|
@ -4,14 +4,58 @@
|
||||
"data_version": "4.0",
|
||||
"CVE_data_meta": {
|
||||
"ID": "CVE-2020-27264",
|
||||
"ASSIGNER": "cve@mitre.org",
|
||||
"STATE": "RESERVED"
|
||||
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
||||
"STATE": "PUBLIC"
|
||||
},
|
||||
"affects": {
|
||||
"vendor": {
|
||||
"vendor_data": [
|
||||
{
|
||||
"vendor_name": "n/a",
|
||||
"product": {
|
||||
"product_data": [
|
||||
{
|
||||
"product_name": "SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A",
|
||||
"version": {
|
||||
"version_data": [
|
||||
{
|
||||
"version_value": "Dana Diabecare RS, AnyDana-i, AnyDana-A All versions prior to 3.0"
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
}
|
||||
]
|
||||
}
|
||||
},
|
||||
"problemtype": {
|
||||
"problemtype_data": [
|
||||
{
|
||||
"description": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "USE OF INSUFFICIENTLY RANDOM VALUES CWE-330"
|
||||
}
|
||||
]
|
||||
}
|
||||
]
|
||||
},
|
||||
"references": {
|
||||
"reference_data": [
|
||||
{
|
||||
"refsource": "MISC",
|
||||
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01",
|
||||
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01"
|
||||
}
|
||||
]
|
||||
},
|
||||
"description": {
|
||||
"description_data": [
|
||||
{
|
||||
"lang": "eng",
|
||||
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
|
||||
"value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications use deterministic keys, which allows unauthenticated, physically proximate attackers to brute-force the keys via Bluetooth Low Energy."
|
||||
}
|
||||
]
|
||||
}
|
||||
|