"-Synchronized-Data."

This commit is contained in:
CVE Team 2020-09-11 13:01:37 +00:00
parent 74216790d3
commit dd5db37e84
No known key found for this signature in database
GPG Key ID: 5708902F06FEF743
7 changed files with 203 additions and 12 deletions

View File

@ -4,14 +4,58 @@
"data_version": "4.0",
"CVE_data_meta": {
"ID": "CVE-2020-16214",
"ASSIGNER": "cve@mitre.org",
"STATE": "RESERVED"
"ASSIGNER": "ics-cert@hq.dhs.gov",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"vendor_name": "n/a",
"product": {
"product_data": [
{
"product_name": "Philips Patient Information Center iX (PICiX), PerformanceBridge Focal Point, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90, IntelliVue X3 and X2.",
"version": {
"version_data": [
{
"version_value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior."
}
]
}
}
]
}
}
]
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "IMPROPER NEUTRALIZATION OF FORMULA ELEMENTS IN A CSV FILE CWE-1236"
}
]
}
]
},
"references": {
"reference_data": [
{
"refsource": "MISC",
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01",
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01"
}
]
},
"description": {
"description_data": [
{
"lang": "eng",
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
"value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software saves user-provided information into a comma-separated value (CSV) file, but it does not neutralize or incorrectly neutralizes special elements that could be interpreted as a command when the file is opened by spreadsheet software."
}
]
}

View File

@ -4,14 +4,58 @@
"data_version": "4.0",
"CVE_data_meta": {
"ID": "CVE-2020-16218",
"ASSIGNER": "cve@mitre.org",
"STATE": "RESERVED"
"ASSIGNER": "ics-cert@hq.dhs.gov",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"vendor_name": "n/a",
"product": {
"product_data": [
{
"product_name": "Philips Patient Information Center iX (PICiX), PerformanceBridge Focal Point, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90, IntelliVue X3 and X2.",
"version": {
"version_data": [
{
"version_value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior."
}
]
}
}
]
}
}
]
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "IMPROPER NEUTRALIZATION OF INPUT DURING WEB PAGE GENERATION ('CROSS-SITE SCRIPTING') CWE-79"
}
]
}
]
},
"references": {
"reference_data": [
{
"refsource": "MISC",
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01",
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01"
}
]
},
"description": {
"description_data": [
{
"lang": "eng",
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
"value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software does not neutralize or incorrectly neutralizes user-controllable input before it is placed in output that is then used as a webpage and served to other users. Successful exploitation could lead to unauthorized access to patient data via a read-only web application."
}
]
}

View File

@ -4,14 +4,58 @@
"data_version": "4.0",
"CVE_data_meta": {
"ID": "CVE-2020-16222",
"ASSIGNER": "cve@mitre.org",
"STATE": "RESERVED"
"ASSIGNER": "ics-cert@hq.dhs.gov",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"vendor_name": "n/a",
"product": {
"product_data": [
{
"product_name": "Philips Patient Information Center iX (PICiX), PerformanceBridge Focal Point, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90, IntelliVue X3 and X2.",
"version": {
"version_data": [
{
"version_value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior."
}
]
}
}
]
}
}
]
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "IMPROPER AUTHENTICATION CWE-287"
}
]
}
]
},
"references": {
"reference_data": [
{
"refsource": "MISC",
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01",
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01"
}
]
},
"description": {
"description_data": [
{
"lang": "eng",
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
"value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. When an actor claims to have a given identity, the software does not prove or insufficiently proves the claim is correct."
}
]
}

View File

@ -4,14 +4,58 @@
"data_version": "4.0",
"CVE_data_meta": {
"ID": "CVE-2020-16228",
"ASSIGNER": "cve@mitre.org",
"STATE": "RESERVED"
"ASSIGNER": "ics-cert@hq.dhs.gov",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"vendor_name": "n/a",
"product": {
"product_data": [
{
"product_name": "Philips Patient Information Center iX (PICiX), PerformanceBridge Focal Point, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90, IntelliVue X3 and X2.",
"version": {
"version_data": [
{
"version_value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior."
}
]
}
}
]
}
}
]
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "IMPROPER CHECK FOR CERTIFICATE REVOCATION CWE-299"
}
]
}
]
},
"references": {
"reference_data": [
{
"refsource": "MISC",
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01",
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-254-01"
}
]
},
"description": {
"description_data": [
{
"lang": "eng",
"value": "** RESERVED ** This candidate has been reserved by an organization or individual that will use it when announcing a new security problem. When the candidate has been publicized, the details for this candidate will be provided."
"value": "Patient Information Center iX (PICiX) Versions B.02, C.02, C.03, PerformanceBridge Focal Point Version A.01, IntelliVue patient monitors MX100, MX400-MX850, and MP2-MP90 Versions N and prior, IntelliVue X3 and X2 Versions N and prior. The software does not check or incorrectly checks the revocation status of a certificate, which may cause it to use a compromised certificate."
}
]
}

View File

@ -71,6 +71,11 @@
"name": "https://www.openssl.org/news/secadv/20200909.txt",
"refsource": "CONFIRM",
"url": "https://www.openssl.org/news/secadv/20200909.txt"
},
{
"refsource": "CONFIRM",
"name": "https://security.netapp.com/advisory/ntap-20200911-0004/",
"url": "https://security.netapp.com/advisory/ntap-20200911-0004/"
}
]
}

View File

@ -66,6 +66,11 @@
"refsource": "GENTOO",
"name": "GLSA-202009-01",
"url": "https://security.gentoo.org/glsa/202009-01"
},
{
"refsource": "CONFIRM",
"name": "https://security.netapp.com/advisory/ntap-20200911-0006/",
"url": "https://security.netapp.com/advisory/ntap-20200911-0006/"
}
]
}

View File

@ -48,6 +48,11 @@
"refsource": "MISC",
"name": "https://www.intel.com/content/www/us/en/security-center/advisory/intel-sa-00404.html",
"url": "https://www.intel.com/content/www/us/en/security-center/advisory/intel-sa-00404.html"
},
{
"refsource": "CONFIRM",
"name": "https://security.netapp.com/advisory/ntap-20200911-0005/",
"url": "https://security.netapp.com/advisory/ntap-20200911-0005/"
}
]
},