{ "data_type": "CVE", "data_format": "MITRE", "data_version": "4.0", "CVE_data_meta": { "ID": "CVE-2020-12037", "ASSIGNER": "ics-cert@hq.dhs.gov", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "vendor_name": "n/a", "product": { "product_data": [ { "product_name": "Baxter PrismaFlex and PrisMax", "version": { "version_data": [ { "version_value": "PrismaFlex all versions, PrisMax all versions prior to 3.x" } ] } } ] } } ] } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "USE OF HARD-CODED PASSWORD CWE-259" } ] } ] }, "references": { "reference_data": [ { "refsource": "MISC", "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01", "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01" } ] }, "description": { "description_data": [ { "lang": "eng", "value": "Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device." } ] } }