{ "data_type": "CVE", "data_format": "MITRE", "data_version": "4.0", "CVE_data_meta": { "ID": "CVE-2020-12035", "ASSIGNER": "ics-cert@hq.dhs.gov", "STATE": "PUBLIC" }, "affects": { "vendor": { "vendor_data": [ { "vendor_name": "n/a", "product": { "product_data": [ { "product_name": "Baxter PrismaFlex and PrisMax", "version": { "version_data": [ { "version_value": "PrismaFlex all versions, PrisMax all versions prior to 3.x" } ] } } ] } } ] } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "IMPROPER AUTHENTICATION CWE-287" } ] } ] }, "references": { "reference_data": [ { "refsource": "MISC", "name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01", "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01" } ] }, "description": { "description_data": [ { "lang": "eng", "value": "Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration." } ] } }