{ "CVE_data_meta": { "ASSIGNER": "ics-cert@hq.dhs.gov", "DATE_PUBLIC": "2022-09-01T17:00:00.000Z", "ID": "CVE-2022-38069", "STATE": "PUBLIC", "TITLE": "Contec Health CMS8000" }, "affects": { "vendor": { "vendor_data": [ { "product": { "product_data": [ { "product_name": "CMS8000 CONTEC ICU CCU Vital Signs Patient Monitor", "version": { "version_data": [ { "version_affected": "=", "version_name": "All", "version_value": "All" } ] } } ] }, "vendor_name": "Contec Health" } ] } }, "credit": [ { "lang": "eng", "value": "Level Nine reported these vulnerabilities to CISA." } ], "data_format": "MITRE", "data_type": "CVE", "data_version": "4.0", "description": { "description_data": [ { "lang": "eng", "value": "Multiple globally default credentials exist across all CMS8000 devices, that once exposed, allow a threat actor with momentary physical access to gain privileged access to any device. Privileged credential access enables the extraction of sensitive patient information or modification of device parameters" } ] }, "generator": { "engine": "Vulnogram 0.0.9" }, "impact": { "cvss": { "attackComplexity": "LOW", "attackVector": "PHYSICAL", "availabilityImpact": "LOW", "baseScore": 4.3, "baseSeverity": "MEDIUM", "confidentialityImpact": "LOW", "integrityImpact": "LOW", "privilegesRequired": "NONE", "scope": "UNCHANGED", "userInteraction": "NONE", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L", "version": "3.1" } }, "problemtype": { "problemtype_data": [ { "description": [ { "lang": "eng", "value": "CWE-798 Use of Hard-coded Credentials" } ] } ] }, "references": { "reference_data": [ { "refsource": "MISC", "url": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01", "name": "https://www.cisa.gov/uscert/ics/advisories/icsma-22-244-01" } ] }, "source": { "advisory": "ICSMA-22-244-01", "discovery": "EXTERNAL" }, "work_around": [ { "lang": "eng", "value": "Contec Health has not responded to requests to work with CISA to mitigate these vulnerabilities. Users of these affected products are invited to contact Contec Health for additional information.\n\nThe following mitigations could assist in reducing the risk for exploitation of vulnerabilities:\n\nDisabling UART functionality at the CPU level\nEnforcing unique device authentication before granting access to the terminal / bootloader\nWhere possible, enforcing secure boot. \nTamper stickers on the device casing to indicate when a device has been opened" } ] }