"CVE_data_meta" : {
"ASSIGNER" : "ics-cert@hq.dhs.gov",
"DATE_PUBLIC" : "2017-08-29T00:00:00",
"ID" : "CVE-2017-12714",
"STATE" : "PUBLIC"
},
"affects" : {
"vendor" : {
"vendor_data" : [
{
"product" : {
"product_data" : [
{
"product_name" : "Accent/Anthem, Accent MRI, Assurity/Allure, and Assurity MRI.",
"version" : {
"version_data" : [
{
"version_value" : "All versions of pacemakers manufactured prior to August 28, 2017"
}
]
}
}
]
},
"vendor_name" : "Abbott Laboratories"
}
]
}
},
"data_format" : "MITRE",
"data_type" : "CVE",
"data_version" : "4.0",
"description" : {
"description_data" : [
{
"lang" : "eng",
"value" : "Abbott Laboratories pacemakers manufactured prior to Aug 28, 2017 do not restrict or limit the number of correctly formatted \"RF wake-up\" commands that can be received, which may allow a nearby attacker to repeatedly send commands to reduce pacemaker battery life. CVSS v3 base score: 5.3, CVSS vector string: AV:A/AC:H/PR:N/UI:N/S:U/C:N/I:N/A:H. Abbott has developed a firmware update to help mitigate the identified vulnerabilities."
}
]
},
"problemtype" : {
"problemtype_data" : [
{
"description" : [
{
"lang" : "eng",
"value" : "Improper Restriction of power consumption CWE-920"
}
]
}
]
},
"references" : {
"reference_data" : [
{
"name" : "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01",
"refsource" : "MISC",
"url" : "https://ics-cert.us-cert.gov/advisories/ICSMA-17-241-01"
},
{
"name" : "100523",
"refsource" : "BID",
"url" : "http://www.securityfocus.com/bid/100523"
}
]
}
}