"data_type": "CVE",
|
|
"data_format": "MITRE",
|
|
"data_version": "4.0",
|
|
"CVE_data_meta": {
|
|
"ID": "CVE-2020-12045",
|
|
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
|
"STATE": "PUBLIC"
|
|
},
|
|
"affects": {
|
|
"vendor": {
|
|
"vendor_data": [
|
|
{
|
|
"vendor_name": "n/a",
|
|
"product": {
|
|
"product_data": [
|
|
{
|
|
"product_name": "Baxter Sigma Spectrum Infusion Pumps",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_value": "Sigma Spectrum v6.x model 35700BAX,Spectrum v8.x model 35700BAX2,Sigma Spectrum v6.x with Wireless Battery Module (WBM) v9,v11,v13,v14,v15,v16,v20D29,v20D30,v20D31,v22D24, Spectrum v8.x w/WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum WBM v17,v20D29,v20D30,v20D31,v22D24,Spectrum LVP v8.x with WBM v17, v20D29,v20D30,v20D31,and v22D24"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"problemtype": {
|
|
"problemtype_data": [
|
|
{
|
|
"description": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "USE OF HARD CODED PASSWORD CWE-259"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"references": {
|
|
"reference_data": [
|
|
{
|
|
"refsource": "MISC",
|
|
"name": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04",
|
|
"url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04"
|
|
}
|
|
]
|
|
},
|
|
"description": {
|
|
"description_data": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "The Baxter Spectrum WBM (v17, v20D29, v20D30, v20D31, and v22D24) when used in conjunction with a Baxter Spectrum v8.x (model 35700BAX2), operates a Telnet service on Port 1023 with hard-coded credentials."
|
|
}
|
|
]
|
|
}
|
|
} |