"CVE_data_meta": {
|
|
"ASSIGNER": "productcert@siemens.com",
|
|
"ID": "CVE-2020-7590",
|
|
"STATE": "PUBLIC"
|
|
},
|
|
"data_format": "MITRE",
|
|
"data_version": "4.0",
|
|
"data_type": "CVE",
|
|
"affects": {
|
|
"vendor": {
|
|
"vendor_data": [
|
|
{
|
|
"vendor_name": "Siemens",
|
|
"product": {
|
|
"product_data": [
|
|
{
|
|
"product_name": "DCA Vantage Analyzer",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_value": "All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"problemtype": {
|
|
"problemtype_data": [
|
|
{
|
|
"description": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "CWE-259: Use of Hard-coded Password"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"description": {
|
|
"description_data": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device."
|
|
}
|
|
]
|
|
},
|
|
"references": {
|
|
"reference_data": [
|
|
{
|
|
"refsource": "MISC",
|
|
"url": "https://www.siemens-healthineers.com/support-documentation/security-advisory",
|
|
"name": "https://www.siemens-healthineers.com/support-documentation/security-advisory"
|
|
}
|
|
]
|
|
}
|
|
} |