mirror of
https://github.com/CVEProject/cvelist.git
synced 2025-07-30 18:04:30 +00:00
113 lines
4.7 KiB
JSON
113 lines
4.7 KiB
JSON
{
|
|
"data_version": "4.0",
|
|
"data_type": "CVE",
|
|
"data_format": "MITRE",
|
|
"CVE_data_meta": {
|
|
"ID": "CVE-2025-0683",
|
|
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
|
"STATE": "PUBLIC"
|
|
},
|
|
"description": {
|
|
"description_data": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
|
|
}
|
|
]
|
|
},
|
|
"problemtype": {
|
|
"problemtype_data": [
|
|
{
|
|
"description": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "CWE-359 Exposure of Private Personal Information to an Unauthorized Actor",
|
|
"cweId": "CWE-359"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"affects": {
|
|
"vendor": {
|
|
"vendor_data": [
|
|
{
|
|
"vendor_name": "Contec Health",
|
|
"product": {
|
|
"product_data": [
|
|
{
|
|
"product_name": "CMS8000 Patient Monitor",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_affected": "=",
|
|
"version_value": "All versions"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"references": {
|
|
"reference_data": [
|
|
{
|
|
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01",
|
|
"refsource": "MISC",
|
|
"name": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
|
|
},
|
|
{
|
|
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication",
|
|
"refsource": "MISC",
|
|
"name": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
|
|
}
|
|
]
|
|
},
|
|
"generator": {
|
|
"engine": "Vulnogram 0.2.0"
|
|
},
|
|
"source": {
|
|
"advisory": "ICSMA-25-030-01",
|
|
"discovery": "EXTERNAL"
|
|
},
|
|
"solution": [
|
|
{
|
|
"lang": "en",
|
|
"supportingMedia": [
|
|
{
|
|
"base64": false,
|
|
"type": "text/html",
|
|
"value": "<p>Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.</p><p>Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's <a target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\">safety communication</a>.</p>\n\n<br>"
|
|
}
|
|
],
|
|
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
|
|
}
|
|
],
|
|
"credits": [
|
|
{
|
|
"lang": "en",
|
|
"value": "An anonymous researcher reported these vulnerabilities to CISA."
|
|
}
|
|
],
|
|
"impact": {
|
|
"cvss": [
|
|
{
|
|
"attackComplexity": "HIGH",
|
|
"attackVector": "NETWORK",
|
|
"availabilityImpact": "NONE",
|
|
"baseScore": 5.9,
|
|
"baseSeverity": "MEDIUM",
|
|
"confidentialityImpact": "HIGH",
|
|
"integrityImpact": "NONE",
|
|
"privilegesRequired": "NONE",
|
|
"scope": "UNCHANGED",
|
|
"userInteraction": "NONE",
|
|
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
|
|
"version": "3.1"
|
|
}
|
|
]
|
|
}
|
|
} |