cvelist/2025/0xxx/CVE-2025-0683.json
2025-01-31 18:00:36 +00:00

113 lines
4.7 KiB
JSON

{
"data_version": "4.0",
"data_type": "CVE",
"data_format": "MITRE",
"CVE_data_meta": {
"ID": "CVE-2025-0683",
"ASSIGNER": "ics-cert@hq.dhs.gov",
"STATE": "PUBLIC"
},
"description": {
"description_data": [
{
"lang": "eng",
"value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
}
]
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-359 Exposure of Private Personal Information to an Unauthorized Actor",
"cweId": "CWE-359"
}
]
}
]
},
"affects": {
"vendor": {
"vendor_data": [
{
"vendor_name": "Contec Health",
"product": {
"product_data": [
{
"product_name": "CMS8000 Patient Monitor",
"version": {
"version_data": [
{
"version_affected": "=",
"version_value": "All versions"
}
]
}
}
]
}
}
]
}
},
"references": {
"reference_data": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01",
"refsource": "MISC",
"name": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication",
"refsource": "MISC",
"name": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication"
}
]
},
"generator": {
"engine": "Vulnogram 0.2.0"
},
"source": {
"advisory": "ICSMA-25-030-01",
"discovery": "EXTERNAL"
},
"solution": [
{
"lang": "en",
"supportingMedia": [
{
"base64": false,
"type": "text/html",
"value": "<p>Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.</p><p>Please note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's <a target=\"_blank\" rel=\"nofollow\" href=\"https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication\">safety communication</a>.</p>\n\n<br>"
}
],
"value": "Per FDA recommendation, CISA recommends users remove any Contec CMS8000 devices from their networks.\n\nPlease note that this device may be re-labeled and sold by resellers. For a list of known re-labeled devices, please refer to FDA's safety communication https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication ."
}
],
"credits": [
{
"lang": "en",
"value": "An anonymous researcher reported these vulnerabilities to CISA."
}
],
"impact": {
"cvss": [
{
"attackComplexity": "HIGH",
"attackVector": "NETWORK",
"availabilityImpact": "NONE",
"baseScore": 5.9,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "HIGH",
"integrityImpact": "NONE",
"privilegesRequired": "NONE",
"scope": "UNCHANGED",
"userInteraction": "NONE",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
"version": "3.1"
}
]
}
}