mirror of
https://github.com/CVEProject/cvelist.git
synced 2025-07-30 18:04:30 +00:00
67 lines
2.2 KiB
JSON
67 lines
2.2 KiB
JSON
{
|
|
"data_type": "CVE",
|
|
"data_format": "MITRE",
|
|
"data_version": "4.0",
|
|
"CVE_data_meta": {
|
|
"ID": "CVE-2019-10962",
|
|
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
|
"STATE": "PUBLIC"
|
|
},
|
|
"affects": {
|
|
"vendor": {
|
|
"vendor_data": [
|
|
{
|
|
"vendor_name": "n/a",
|
|
"product": {
|
|
"product_data": [
|
|
{
|
|
"product_name": "BD Alaris Gateway Workstation",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_value": "versions 1.0.13,1.1.3 Build 10,1.1.3 MR Build 11,1.1.5,1.1.6"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"problemtype": {
|
|
"problemtype_data": [
|
|
{
|
|
"description": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "IMPROPER ACCESS CONTROL CWE-284"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"references": {
|
|
"reference_data": [
|
|
{
|
|
"refsource": "MISC",
|
|
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-164-01",
|
|
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-164-01"
|
|
},
|
|
{
|
|
"refsource": "BID",
|
|
"name": "108763",
|
|
"url": "http://www.securityfocus.com/bid/108763"
|
|
}
|
|
]
|
|
},
|
|
"description": {
|
|
"description_data": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "BD Alaris Gateway versions, 1.0.13,1.1.3 Build 10,1.1.3 MR Build 11,1.1.5, and 1.1.6, The web browser user interface on the Alaris Gateway Workstation does not prevent an attacker with knowledge of the IP address of the Alaris Gateway Workstation terminal to gain access to the status and configuration information of the device."
|
|
}
|
|
]
|
|
}
|
|
} |