"description_data": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "Insufficiently Protected Credentials vulnerability in Baxter Welch Allyn Configuration Tool may allow Remote Services with Stolen Credentials.This issue affects Welch Allyn Configuration Tool: versions 1.9.4.1 and prior."
|
|
}
|
|
]
|
|
},
|
|
"problemtype": {
|
|
"problemtype_data": [
|
|
{
|
|
"description": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "CWE-522 Insufficiently Protected Credentials",
|
|
"cweId": "CWE-522"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"affects": {
|
|
"vendor": {
|
|
"vendor_data": [
|
|
{
|
|
"vendor_name": "Baxter",
|
|
"product": {
|
|
"product_data": [
|
|
{
|
|
"product_name": "Welch Allyn Configuration Tool",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_affected": "<=",
|
|
"version_name": "0",
|
|
"version_value": "1.9.4.1"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"references": {
|
|
"reference_data": [
|
|
{
|
|
"url": "https://cisa.gov/news-events/ics-medical-advisories/icsma-24-151-01",
|
|
"refsource": "MISC",
|
|
"name": "https://cisa.gov/news-events/ics-medical-advisories/icsma-24-151-01"
|
|
}
|
|
]
|
|
},
|
|
"generator": {
|
|
"engine": "Vulnogram 0.2.0"
|
|
},
|
|
"source": {
|
|
"advisory": "ICSMA-24-151-01",
|
|
"discovery": "UNKNOWN"
|
|
},
|
|
"solution": [
|
|
{
|
|
"lang": "en",
|
|
"supportingMedia": [
|
|
{
|
|
"base64": false,
|
|
"type": "text/html",
|
|
"value": "<p>Baxter has found no evidence to date of any compromise of personal or health data. Baxter will release a software update for all impacted software to address this vulnerability. A new version of the product that mitigates the vulnerability will be available as follows:</p><ul><li>Welch Allyn Product Configuration Tool versions 1.9.4.2: Available Q3 2024</li><li>No user action will be required once the update is released.</li></ul><p>Baxter recommends the following workarounds to help reduce risk:</p><ul><li>Apply proper network and physical security controls.</li><li>The Welch Allyn Configuration Tool has been removed from public access. Customers are advised to contact Baxter Technical Support or their Baxter Project Manager to create configuration files, as needed. Baxter Technical Support can be reached at (800)535-6663, option 2.</li></ul>"
|
|
}
|
|
],
|
|
"value": "Baxter has found no evidence to date of any compromise of personal or health data. Baxter will release a software update for all impacted software to address this vulnerability. A new version of the product that mitigates the vulnerability will be available as follows:\n\n * Welch Allyn Product Configuration Tool versions 1.9.4.2: Available Q3 2024\n * No user action will be required once the update is released.\n\n\nBaxter recommends the following workarounds to help reduce risk:\n\n * Apply proper network and physical security controls.\n * The Welch Allyn Configuration Tool has been removed from public access. Customers are advised to contact Baxter Technical Support or their Baxter Project Manager to create configuration files, as needed. Baxter Technical Support can be reached at (800)535-6663, option 2."
|
|
}
|
|
],
|
|
"credits": [
|
|
{
|
|
"lang": "en",
|
|
"value": "Baxter reported this vulnerability to CISA."
|
|
}
|
|
]
|
|
} |