"CVE_data_meta" : {
"ASSIGNER" : "ics-cert@hq.dhs.gov",
"DATE_PUBLIC" : "2018-08-08T00:00:00",
"ID" : "CVE-2018-14781",
"STATE" : "PUBLIC"
},
"affects" : {
"vendor" : {
"vendor_data" : [
{
"product" : {
"product_data" : [
{
"product_name" : "Medtronic insulin pump",
"version" : {
"version_data" : [
{
"version_value" : "MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G"
}
]
}
}
]
},
"vendor_name" : "ICS-CERT"
}
]
}
},
"data_format" : "MITRE",
"data_type" : "CVE",
"data_version" : "4.0",
"description" : {
"description_data" : [
{
"lang" : "eng",
"value" : "Medtronic MMT 508 MiniMed insulin pump, 522 / MMT - 722 Paradigm REAL-TIME, 523 / MMT - 723 Paradigm Revel, 523K / MMT - 723K Paradigm Revel, and 551 / MMT - 751 MiniMed 530G The models identified above, when paired with a remote controller and having the \"easy bolus\" and \"remote bolus\" options enabled (non-default), are vulnerable to a capture-replay attack. An attacker can capture the wireless transmissions between the remote controller and the pump and replay them to cause an insulin (bolus) delivery."
}
]
},
"problemtype" : {
"problemtype_data" : [
{
"description" : [
{
"lang" : "eng",
"value" : "AUTHENTICATION BYPASS BY CAPTURE-REPLAY CWE-294"
}
]
}
]
},
"references" : {
"reference_data" : [
{
"name" : "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02",
"refsource" : "MISC",
"url" : "https://ics-cert.us-cert.gov/advisories/ICSMA-18-219-02"
},
{
"name" : "105044",
"refsource" : "BID",
"url" : "http://www.securityfocus.com/bid/105044"
}
]
}
}