"CVE_data_meta": {
|
|
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
|
"ID": "CVE-2021-43355",
|
|
"STATE": "PUBLIC",
|
|
"TITLE": "Fresenius Kabi Agilia Connect Infusion System use of client side authentication"
|
|
},
|
|
"affects": {
|
|
"vendor": {
|
|
"vendor_data": [
|
|
{
|
|
"product": {
|
|
"product_data": [
|
|
{
|
|
"product_name": "Vigilant Software Suite (Mastermed Dashboard)",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_affected": "<",
|
|
"version_value": "2.0.1.3"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
},
|
|
"vendor_name": "Fresenius Kabi"
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"credit": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "Julian Suleder (ERNW Research GmbH), Nils Emmerich (ERNW Research GmbH), Raphael Pavlidis (ERNW Research GmbH), and Dr. Oliver Matula (ERNW Enno Rey Netzwerke GmbH) reported these vulnerabilities to the German Federal Office for Information Security (BSI) in the context of the BSI project ManiMed (Medical Device Manipulation Project)."
|
|
}
|
|
],
|
|
"data_format": "MITRE",
|
|
"data_type": "CVE",
|
|
"data_version": "4.0",
|
|
"description": {
|
|
"description_data": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "Fresenius Kabi Vigilant Software Suite (Mastermed Dashboard) version 2.0.1.3 allows user input to be validated on the client side without authentication by the server. The server should not rely on the correctness of the data because users might not support or block JavaScript or intentionally bypass the client-side checks. An attacker with knowledge of the service user could circumvent the client-side control and login with service privileges."
|
|
}
|
|
]
|
|
},
|
|
"generator": {
|
|
"engine": "Vulnogram 0.0.9"
|
|
},
|
|
"impact": {
|
|
"cvss": {
|
|
"attackComplexity": "LOW",
|
|
"attackVector": "NETWORK",
|
|
"availabilityImpact": "LOW",
|
|
"baseScore": 7.3,
|
|
"baseSeverity": "HIGH",
|
|
"confidentialityImpact": "LOW",
|
|
"integrityImpact": "LOW",
|
|
"privilegesRequired": "NONE",
|
|
"scope": "UNCHANGED",
|
|
"userInteraction": "NONE",
|
|
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:L/I:L/A:L",
|
|
"version": "3.1"
|
|
}
|
|
},
|
|
"problemtype": {
|
|
"problemtype_data": [
|
|
{
|
|
"description": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "CWE-603 Use of client side authentication"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"references": {
|
|
"reference_data": [
|
|
{
|
|
"refsource": "MISC",
|
|
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01",
|
|
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-21-355-01"
|
|
}
|
|
]
|
|
},
|
|
"solution": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "Fresenius Kabi has created new versions to address these vulnerabilities:\n\u2022\tLink+ v3.0 (D16 or later)\n\u2022\tVSS v1.0.3 (or later)\n\u2022\tAgilia Connect Pumps Wifi Module (D29 or later)\n\u2022\tAgilia Connect Partner v3.3.2 (or later) \nFresenius Kabi has initiated communication on this topic in April 2021 with customers to inform them about availability of the new versions in corresponding countries. Contact Fresenius Kabi online or by phone at 1-800-333-6925 for more information on updating to the newest version.\nFresenius Kabi also identified that early Link+ devices (approximatively 1200 devices) would need hardware change to support D16 or later firmware. Until those devices can be replaced in customers\u2019 installation, Fresenius Kabi recommends users to rely on CISA recommendation. \n"
|
|
}
|
|
],
|
|
"source": {
|
|
"advisory": "ICSMA-21-355-01",
|
|
"discovery": "EXTERNAL"
|
|
},
|
|
"work_around": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "CISA recommends users take defensive measures to minimize the risk of exploitation of these vulnerabilities. Specifically, users should:\n\u2022\tMinimize network exposure for all control system devices and/or systems, and ensure they are not accessible from the Internet.\n\u2022\tLocate control system networks and remote devices behind firewalls and isolate them from the business network.\n\u2022\tWhen remote access is required, use secure methods, such as virtual private networks (VPNs), recognizing VPNs may have vulnerabilities and should be updated to the most current version available. Also recognize VPN is only as secure as its connected devices.\n"
|
|
}
|
|
]
|
|
} |