mirror of
https://github.com/CVEProject/cvelist.git
synced 2025-07-29 05:56:59 +00:00
143 lines
6.4 KiB
JSON
143 lines
6.4 KiB
JSON
{
|
|
"CVE_data_meta": {
|
|
"ASSIGNER": "ics-cert@hq.dhs.gov",
|
|
"ID": "CVE-2020-25150",
|
|
"STATE": "PUBLIC",
|
|
"TITLE": "B. Braun SpaceCom, Battery Pack SP with Wi-Fi, and Data module compactplus"
|
|
},
|
|
"affects": {
|
|
"vendor": {
|
|
"vendor_data": [
|
|
{
|
|
"product": {
|
|
"product_data": [
|
|
{
|
|
"product_name": "SpaceCom",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_affected": "<=",
|
|
"version_value": "U61"
|
|
},
|
|
{
|
|
"version_affected": "<=",
|
|
"version_value": "L81"
|
|
}
|
|
]
|
|
}
|
|
},
|
|
{
|
|
"product_name": "Battery pack with Wi-Fi",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_affected": "<=",
|
|
"version_value": "U61"
|
|
},
|
|
{
|
|
"version_affected": "<=",
|
|
"version_value": "L81"
|
|
}
|
|
]
|
|
}
|
|
},
|
|
{
|
|
"product_name": "Data module compactplus",
|
|
"version": {
|
|
"version_data": [
|
|
{
|
|
"version_affected": "=",
|
|
"version_value": "A10"
|
|
},
|
|
{
|
|
"version_affected": "=",
|
|
"version_value": "A11"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
},
|
|
"vendor_name": "B. Braun Melsungen AG"
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"credit": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "Julian Suleder, Nils Emmerich, and Birk Kauer of ERNW Research GmbH; Dr. Oliver Matula of ERNW Enno Rey Netzwerke GmbH, reported these vulnerabilities to the Federal Office for Information Security (BSI), Germany, in the context of the BSI project ManiMed (Manipulation of medical devices)."
|
|
}
|
|
],
|
|
"data_format": "MITRE",
|
|
"data_type": "CVE",
|
|
"data_version": "4.0",
|
|
"description": {
|
|
"description_data": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "A relative path traversal attack in the B. Braun Melsungen AG SpaceCom Version L81/U61 and earlier, and the Data module compactplus Versions A10 and A11 allows attackers with service user privileges to upload arbitrary files. By uploading a specially crafted tar file an attacker can execute arbitrary commands."
|
|
}
|
|
]
|
|
},
|
|
"generator": {
|
|
"engine": "Vulnogram 0.0.9"
|
|
},
|
|
"impact": {
|
|
"cvss": {
|
|
"attackComplexity": "LOW",
|
|
"attackVector": "NETWORK",
|
|
"availabilityImpact": "LOW",
|
|
"baseScore": 7.6,
|
|
"baseSeverity": "HIGH",
|
|
"confidentialityImpact": "LOW",
|
|
"integrityImpact": "HIGH",
|
|
"privilegesRequired": "LOW",
|
|
"scope": "UNCHANGED",
|
|
"userInteraction": "NONE",
|
|
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:L/UI:N/S:U/C:L/I:H/A:L",
|
|
"version": "3.1"
|
|
}
|
|
},
|
|
"problemtype": {
|
|
"problemtype_data": [
|
|
{
|
|
"description": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "CWE-23: Relative Path Traversal"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"references": {
|
|
"reference_data": [
|
|
{
|
|
"name": "https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02",
|
|
"refsource": "CONFIRM",
|
|
"url": "https://www.cisa.gov/uscert/ics/advisories/icsma-20-296-02"
|
|
},
|
|
{
|
|
"name": "https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html",
|
|
"refsource": "CONFIRM",
|
|
"url": "https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html"
|
|
}
|
|
]
|
|
},
|
|
"solution": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "B. Braun recommends applying updates:\n\n SpaceCom: Version U62 or later (United States), L82 or later (outside the United States)\n Battery Pack SP with Wi-Fi: Version U62 or later (United States), L82 or later (outside the United States)\n Data module compactplus: Version A12 or later\n\nPlease contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html\n"
|
|
}
|
|
],
|
|
"source": {
|
|
"discovery": "EXTERNAL"
|
|
},
|
|
"work_around": [
|
|
{
|
|
"lang": "eng",
|
|
"value": "As a general security measure, B. Braun recommends protecting the network with appropriate mechanisms:\n\n Ensure the devices are not accessible directly from the Internet.\n Use a firewall and isolate the medical devices from the business network. \n\nPlease contact your local B. Braun organization to request further help. For more information please see the B. Braun Security Advisory. https://www.bbraun.com/en/products-and-therapies/services/b-braun-vulnerability-disclosure-policy/security-advisory.html"
|
|
}
|
|
]
|
|
} |