"CVE_data_meta" : {
|
|
"ASSIGNER" : "ics-cert@hq.dhs.gov",
|
|
"DATE_PUBLIC" : "2018-02-27T00:00:00",
|
|
"ID" : "CVE-2018-5446",
|
|
"STATE" : "PUBLIC"
|
|
},
|
|
"affects" : {
|
|
"vendor" : {
|
|
"vendor_data" : [
|
|
{
|
|
"product" : {
|
|
"product_data" : [
|
|
{
|
|
"product_name" : "Medtronic 2090 CareLink Programmer",
|
|
"version" : {
|
|
"version_data" : [
|
|
{
|
|
"version_value" : "All versions are affected"
|
|
}
|
|
]
|
|
}
|
|
}
|
|
]
|
|
},
|
|
"vendor_name" : "ICS-CERT"
|
|
}
|
|
]
|
|
}
|
|
},
|
|
"data_format" : "MITRE",
|
|
"data_type" : "CVE",
|
|
"data_version" : "4.0",
|
|
"description" : {
|
|
"description_data" : [
|
|
{
|
|
"lang" : "eng",
|
|
"value" : "All versions of the Medtronic 2090 Carelink Programmer are affected by a per-product username and password that is stored in a recoverable format which could allow an attacker with physical access to a 2090 Programmer to obtain per-product credentials to the software deployment network."
|
|
}
|
|
]
|
|
},
|
|
"problemtype" : {
|
|
"problemtype_data" : [
|
|
{
|
|
"description" : [
|
|
{
|
|
"lang" : "eng",
|
|
"value" : "STORING PASSWORDS IN A RECOVERABLE FORMAT CWE-257"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
"references" : {
|
|
"reference_data" : [
|
|
{
|
|
"name" : "https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01",
|
|
"refsource" : "MISC",
|
|
"url" : "https://ics-cert.us-cert.gov/advisories/ICSMA-18-058-01"
|
|
}
|
|
]
|
|
}
|
|
}
|