cvelist/2019/6xxx/CVE-2019-6538.json
2019-03-26 14:00:44 +00:00

124 lines
5.8 KiB
JSON

{
"data_type": "CVE",
"data_format": "MITRE",
"data_version": "4.0",
"CVE_data_meta": {
"ID": "CVE-2019-6538",
"ASSIGNER": "ics-cert@hq.dhs.gov",
"STATE": "PUBLIC"
},
"affects": {
"vendor": {
"vendor_data": [
{
"vendor_name": "Medtronic",
"product": {
"product_data": [
{
"product_name": "Medtronic Conexus Radio Frequency Telemetry Protocol",
"version": {
"version_data": [
{
"version_value": "MyCareLink Monitor versions 24950 and 24952"
},
{
"version_value": "CareLink Monitor version 2490C"
},
{
"version_value": "CareLink 2090 Programmer"
},
{
"version_value": "Amplia CRT-D"
},
{
"version_value": "Claria CRT-D"
},
{
"version_value": "Compia CRT-D"
},
{
"version_value": "Concerto CRT-D"
},
{
"version_value": "Concerto II CRT-D"
},
{
"version_value": "Consulta CRT-D"
},
{
"version_value": "Evera ICD"
},
{
"version_value": "Maximo II CRT-D and ICD"
},
{
"version_value": "Mirro ICD"
},
{
"version_value": "Nayamed ND ICD"
},
{
"version_value": "Primo ICD"
},
{
"version_value": "Protecta ICD and CRT-D"
},
{
"version_value": "Secura ICD"
},
{
"version_value": "Virtuoso ICD"
},
{
"version_value": "Virtuoso II ICD"
},
{
"version_value": "Visia AF ICD"
},
{
"version_value": "Viva CRT-D"
}
]
}
}
]
}
}
]
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "Improper access control CWE-284"
}
]
}
]
},
"references": {
"reference_data": [
{
"refsource": "BID",
"name": "107544",
"url": "http://www.securityfocus.com/bid/107544"
},
{
"refsource": "CONFIRM",
"name": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01",
"url": "https://ics-cert.us-cert.gov/advisories/ICSMA-19-080-01"
}
]
},
"description": {
"description_data": [
{
"lang": "eng",
"value": "The Conexus telemetry protocol utilized within Medtronic MyCareLink Monitor versions 24950 and 24952, CareLink Monitor version 2490C, CareLink 2090 Programmer, Amplia CRT-D, Claria CRT-D, Compia CRT-D, Concerto CRT-D, Concerto II CRT-D, Consulta CRT-D, Evera ICD, Maximo II CRT-D and ICD, Mirro ICD, Nayamed ND ICD, Primo ICD, Protecta ICD and CRT-D, Secura ICD, Virtuoso ICD, Virtuoso II ICD, Visia AF ICD, and Viva CRT-D does not implement authentication or authorization. An attacker with adjacent short-range access to an affected product, in situations where the product\u2019s radio is turned on, can inject, replay, modify, and/or intercept data within the telemetry communication. This communication protocol provides the ability to read and write memory values to affected implanted cardiac devices; therefore, an attacker could exploit this communication protocol to change memory in the implanted cardiac device."
}
]
}
}