"id": "CVE-2024-48974",
|
||
|
"sourceIdentifier": "productsecurity@baxter.com",
|
||
|
"published": "2024-11-14T22:15:18.727",
|
||
|
"lastModified": "2024-11-14T22:15:18.727",
|
||
|
"vulnStatus": "Received",
|
||
|
"cveTags": [],
|
||
|
"descriptions": [
|
||
|
{
|
||
|
"lang": "en",
|
||
|
"value": "The ventilator does not perform proper file integrity checks when adopting firmware updates. This makes it possible for an attacker to force unauthorized changes to the device's configuration settings and/or compromise device functionality by pushing a compromised/illegitimate firmware file. This could disrupt the function of the device and/or cause unauthorized information disclosure."
|
||
|
}
|
||
|
],
|
||
|
"metrics": {
|
||
|
"cvssMetricV31": [
|
||
|
{
|
||
|
"source": "productsecurity@baxter.com",
|
||
|
"type": "Secondary",
|
||
|
"cvssData": {
|
||
|
"version": "3.1",
|
||
|
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
|
||
|
"attackVector": "LOCAL",
|
||
|
"attackComplexity": "LOW",
|
||
|
"privilegesRequired": "NONE",
|
||
|
"userInteraction": "NONE",
|
||
|
"scope": "CHANGED",
|
||
|
"confidentialityImpact": "HIGH",
|
||
|
"integrityImpact": "HIGH",
|
||
|
"availabilityImpact": "HIGH",
|
||
|
"baseScore": 9.3,
|
||
|
"baseSeverity": "CRITICAL"
|
||
|
},
|
||
|
"exploitabilityScore": 2.5,
|
||
|
"impactScore": 6.0
|
||
|
}
|
||
|
]
|
||
|
},
|
||
|
"weaknesses": [
|
||
|
{
|
||
|
"source": "productsecurity@baxter.com",
|
||
|
"type": "Secondary",
|
||
|
"description": [
|
||
|
{
|
||
|
"lang": "en",
|
||
|
"value": "CWE-494"
|
||
|
}
|
||
|
]
|
||
|
}
|
||
|
],
|
||
|
"references": [
|
||
|
{
|
||
|
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01",
|
||
|
"source": "productsecurity@baxter.com"
|
||
|
}
|
||
|
]
|
||
|
}
|