2023-04-24 12:24:31 +02:00
{
"id" : "CVE-2015-3952" ,
"sourceIdentifier" : "ics-cert@hq.dhs.gov" ,
"published" : "2019-03-25T16:29:00.303" ,
"lastModified" : "2019-10-09T23:14:05.410" ,
"vulnStatus" : "Modified" ,
2024-07-14 02:06:08 +00:00
"cveTags" : [ ] ,
2023-04-24 12:24:31 +02:00
"descriptions" : [
{
"lang" : "en" ,
"value" : "Wireless keys are stored in plain text on Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior. Hospira recommends that customers close Port 20/FTP and Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
} ,
{
"lang" : "es" ,
"value" : "Las claves inal\u00e1mbricas se almacenan como texto en claro en Hospira Plum A+ Infusion System en su versi\u00f3n 13.4 y anteriores, Plum A+3 Infusion System en su versi\u00f3n 13.6 y anteriores y Symbiq Infusion System, en su versi\u00f3n 3.13 y anteriores. Hospira recomienda con los usuarios cierren el Puerto 20/FTP y el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado tambi\u00e9n el Plum 360 Infusion System que no es vulnerable a este problema."
}
] ,
"metrics" : {
"cvssMetricV30" : [
{
"source" : "nvd@nist.gov" ,
"type" : "Primary" ,
"cvssData" : {
"version" : "3.0" ,
"vectorString" : "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N" ,
"attackVector" : "NETWORK" ,
"attackComplexity" : "LOW" ,
"privilegesRequired" : "NONE" ,
"userInteraction" : "NONE" ,
"scope" : "UNCHANGED" ,
"confidentialityImpact" : "HIGH" ,
"integrityImpact" : "NONE" ,
"availabilityImpact" : "NONE" ,
"baseScore" : 7.5 ,
"baseSeverity" : "HIGH"
} ,
"exploitabilityScore" : 3.9 ,
"impactScore" : 3.6
}
] ,
"cvssMetricV2" : [
{
"source" : "nvd@nist.gov" ,
"type" : "Primary" ,
"cvssData" : {
"version" : "2.0" ,
"vectorString" : "AV:N/AC:L/Au:N/C:P/I:N/A:N" ,
"accessVector" : "NETWORK" ,
"accessComplexity" : "LOW" ,
"authentication" : "NONE" ,
"confidentialityImpact" : "PARTIAL" ,
"integrityImpact" : "NONE" ,
"availabilityImpact" : "NONE" ,
"baseScore" : 5.0
} ,
"baseSeverity" : "MEDIUM" ,
"exploitabilityScore" : 10.0 ,
"impactScore" : 2.9 ,
"acInsufInfo" : false ,
"obtainAllPrivilege" : false ,
"obtainUserPrivilege" : false ,
"obtainOtherPrivilege" : false ,
"userInteractionRequired" : false
}
]
} ,
"weaknesses" : [
{
"source" : "nvd@nist.gov" ,
"type" : "Primary" ,
"description" : [
{
"lang" : "en" ,
"value" : "CWE-200"
}
]
} ,
{
"source" : "ics-cert@hq.dhs.gov" ,
"type" : "Secondary" ,
"description" : [
{
"lang" : "en" ,
"value" : "CWE-312"
}
]
}
] ,
"configurations" : [
{
"operator" : "AND" ,
"nodes" : [
{
"operator" : "OR" ,
"negate" : false ,
"cpeMatch" : [
{
"vulnerable" : true ,
"criteria" : "cpe:2.3:o:pifzer:plum_a\\+_infusion_system_firmware:*:*:*:*:*:*:*:*" ,
"versionEndIncluding" : "13.4" ,
"matchCriteriaId" : "61C1A9DD-F143-4D0C-871C-B6CD7AF9DAB2"
}
]
} ,
{
"operator" : "OR" ,
"negate" : false ,
"cpeMatch" : [
{
"vulnerable" : false ,
"criteria" : "cpe:2.3:h:pifzer:plum_a\\+_infusion_system:-:*:*:*:*:*:*:*" ,
"matchCriteriaId" : "1DB2490B-0318-4770-BF45-CD7527F15D7F"
}
]
}
]
} ,
{
"operator" : "AND" ,
"nodes" : [
{
"operator" : "OR" ,
"negate" : false ,
"cpeMatch" : [
{
"vulnerable" : true ,
"criteria" : "cpe:2.3:o:pifzer:plum_a\\+3_infusion_system_firmware:*:*:*:*:*:*:*:*" ,
"versionEndIncluding" : "13.6" ,
"matchCriteriaId" : "8536E705-89E8-47CB-9567-6AD65FBA0F1B"
}
]
} ,
{
"operator" : "OR" ,
"negate" : false ,
"cpeMatch" : [
{
"vulnerable" : false ,
"criteria" : "cpe:2.3:h:pifzer:plum_a\\+3_infusion_system:-:*:*:*:*:*:*:*" ,
"matchCriteriaId" : "423AA561-8E38-4378-814B-1008B96F27A6"
}
]
}
]
} ,
{
"operator" : "AND" ,
"nodes" : [
{
"operator" : "OR" ,
"negate" : false ,
"cpeMatch" : [
{
"vulnerable" : true ,
"criteria" : "cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*" ,
"versionEndIncluding" : "3.13" ,
"matchCriteriaId" : "C6DD5141-72AB-4694-8595-B4BED8EC7773"
}
]
} ,
{
"operator" : "OR" ,
"negate" : false ,
"cpeMatch" : [
{
"vulnerable" : false ,
"criteria" : "cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*" ,
"matchCriteriaId" : "5C295E1A-BF60-476D-B972-5C5C28D7633B"
}
]
}
]
}
] ,
"references" : [
{
"url" : "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01" ,
"source" : "ics-cert@hq.dhs.gov" ,
"tags" : [
"Mitigation" ,
"US Government Resource" ,
"Third Party Advisory"
]
}
]
}