{ "id": "CVE-2021-33886", "sourceIdentifier": "cve@mitre.org", "published": "2021-08-25T12:15:17.130", "lastModified": "2022-07-12T17:42:04.277", "vulnStatus": "Analyzed", "descriptions": [ { "lang": "en", "value": "An improper sanitization of input vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote unauthenticated attacker to gain user-level command-line access by passing a raw external string straight through to printf statements. The attacker is required to be on the same network as the device." }, { "lang": "es", "value": "Una vulnerabilidad de saneamiento inapropiado de la entrada en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite que un atacante remoto no autenticado obtenga acceso a la l\u00ednea de comandos a nivel de usuario pasando una cadena externa sin procesar directamente a las declaraciones printf. El atacante debe estar en la misma red que el dispositivo." } ], "metrics": { "cvssMetricV31": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H", "attackVector": "ADJACENT_NETWORK", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "UNCHANGED", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "availabilityImpact": "HIGH", "baseScore": 8.8, "baseSeverity": "HIGH" }, "exploitabilityScore": 2.8, "impactScore": 5.9 }, { "source": "cve@mitre.org", "type": "Secondary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:N", "attackVector": "ADJACENT_NETWORK", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "UNCHANGED", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "availabilityImpact": "NONE", "baseScore": 8.1, "baseSeverity": "HIGH" }, "exploitabilityScore": 2.8, "impactScore": 5.2 } ], "cvssMetricV2": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "2.0", "vectorString": "AV:A/AC:L/Au:N/C:P/I:P/A:P", "accessVector": "ADJACENT_NETWORK", "accessComplexity": "LOW", "authentication": "NONE", "confidentialityImpact": "PARTIAL", "integrityImpact": "PARTIAL", "availabilityImpact": "PARTIAL", "baseScore": 5.8 }, "baseSeverity": "MEDIUM", "exploitabilityScore": 6.5, "impactScore": 6.4, "acInsufInfo": false, "obtainAllPrivilege": false, "obtainUserPrivilege": false, "obtainOtherPrivilege": false, "userInteractionRequired": false } ] }, "weaknesses": [ { "source": "nvd@nist.gov", "type": "Primary", "description": [ { "lang": "en", "value": "CWE-134" } ] } ], "configurations": [ { "operator": "AND", "nodes": [ { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": true, "criteria": "cpe:2.3:o:bbraun:spacecom2:*:*:*:*:*:*:*:*", "versionEndExcluding": "012u000062", "matchCriteriaId": "C2D8F2AC-5506-405C-857A-EE060C209A89" } ] }, { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": false, "criteria": "cpe:2.3:h:bbraun:infusomat_large_volume_pump_871305u:-:*:*:*:*:*:*:*", "matchCriteriaId": "A6C9420B-B015-4079-8B60-EF9C9C058193" }, { "vulnerable": false, "criteria": "cpe:2.3:h:bbraun:spacestation_8713142u:-:*:*:*:*:*:*:*", "matchCriteriaId": "D6B64426-0B44-4B68-A80A-2BB6D04B3F88" } ] } ] } ], "references": [ { "url": "https://www.bbraunusa.com/en.htm", "source": "cve@mitre.org", "tags": [ "Broken Link" ] }, { "url": "https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/", "source": "cve@mitre.org", "tags": [ "Exploit", "Third Party Advisory" ] } ] }