{ "id": "CVE-2020-27258", "sourceIdentifier": "ics-cert@hq.dhs.gov", "published": "2021-01-19T21:15:14.047", "lastModified": "2021-01-22T19:51:14.890", "vulnStatus": "Analyzed", "cveTags": [], "descriptions": [ { "lang": "en", "value": "In SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i and AnyDana-A, an information disclosure vulnerability in the communication protocol of the insulin pump and its AnyDana-i and AnyDana-A mobile applications allows unauthenticated attackers to extract the pump\u2019s keypad lock PIN via Bluetooth Low Energy." }, { "lang": "es", "value": "En SOOIL Developments Co., Ltd Diabecare RS, AnyDana-i y AnyDana-A, una vulnerabilidad de divulgaci\u00f3n de informaci\u00f3n en el protocolo de comunicaci\u00f3n de la bomba de insulina y sus aplicaciones m\u00f3viles AnyDana-i y AnyDana-A permite a atacantes no autenticados extraer el bloqueo del teclado de la bomba. PIN por medio de Bluetooth Low Energy" } ], "metrics": { "cvssMetricV31": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N", "attackVector": "ADJACENT_NETWORK", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "UNCHANGED", "confidentialityImpact": "HIGH", "integrityImpact": "NONE", "availabilityImpact": "NONE", "baseScore": 6.5, "baseSeverity": "MEDIUM" }, "exploitabilityScore": 2.8, "impactScore": 3.6 } ], "cvssMetricV2": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "2.0", "vectorString": "AV:A/AC:L/Au:N/C:P/I:N/A:N", "accessVector": "ADJACENT_NETWORK", "accessComplexity": "LOW", "authentication": "NONE", "confidentialityImpact": "PARTIAL", "integrityImpact": "NONE", "availabilityImpact": "NONE", "baseScore": 3.3 }, "baseSeverity": "LOW", "exploitabilityScore": 6.5, "impactScore": 2.9, "acInsufInfo": false, "obtainAllPrivilege": false, "obtainUserPrivilege": false, "obtainOtherPrivilege": false, "userInteractionRequired": false } ] }, "weaknesses": [ { "source": "nvd@nist.gov", "type": "Primary", "description": [ { "lang": "en", "value": "CWE-522" } ] }, { "source": "ics-cert@hq.dhs.gov", "type": "Secondary", "description": [ { "lang": "en", "value": "CWE-522" } ] } ], "configurations": [ { "nodes": [ { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": true, "criteria": "cpe:2.3:a:sooil:anydana-a:*:*:*:*:*:*:*:*", "versionEndExcluding": "3.0", "matchCriteriaId": "712D6D9A-8A9A-412A-94B0-54DCD142C5B1" }, { "vulnerable": true, "criteria": "cpe:2.3:a:sooil:anydana-i:*:*:*:*:*:*:*:*", "versionEndExcluding": "3.0", "matchCriteriaId": "0C1AA992-E68C-4A87-8C9A-F487376D8825" } ] } ] }, { "operator": "AND", "nodes": [ { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": false, "criteria": "cpe:2.3:h:sooil:dana_diabecare_rs:-:*:*:*:*:*:*:*", "matchCriteriaId": "84960250-D074-4153-902D-C8C27A948076" } ] }, { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": true, "criteria": "cpe:2.3:o:sooil:dana_diabecare_rs_firmware:*:*:*:*:*:*:*:*", "versionEndExcluding": "3.0", "matchCriteriaId": "110D70B7-66A1-4DF0-B080-F0C95FA98B3F" } ] } ] } ], "references": [ { "url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-012-01", "source": "ics-cert@hq.dhs.gov", "tags": [ "Third Party Advisory", "US Government Resource" ] } ] }