{ "id": "CVE-2020-7590", "sourceIdentifier": "productcert@siemens.com", "published": "2020-10-13T16:15:21.780", "lastModified": "2024-11-21T05:37:25.887", "vulnStatus": "Modified", "cveTags": [], "descriptions": [ { "lang": "en", "value": "A vulnerability has been identified in DCA Vantage Analyzer (All versions < V4.5 are affected by CVE-2020-7590. In addition, serial numbers < 40000 running software V4.4.0 are also affected by CVE-2020-15797). Affected devices use a hard-coded password to protect the onboard database. This could allow an attacker to read and or modify the onboard database. Successful exploitation requires direct physical access to the device." }, { "lang": "es", "value": "Se ha identificado una vulnerabilidad en DCA Vantage Analyzer (todas las versiones anteriores a V4.5 est\u00e1n afectadas por CVE-2020-7590. Adem\u00e1s, los n\u00fameros de serial anteriores a 40000 que ejecutan el software versi\u00f3n V4.4.0 tambi\u00e9n est\u00e1n afectadas por CVE-2020-15797). Los dispositivos afectados usan una contrase\u00f1a embebida para proteger la base de datos integrada. Esto podr\u00eda permitir a un atacante leer o modificar la base de datos integrada. Una explotaci\u00f3n con \u00e9xito requiere acceso f\u00edsico directo al dispositivo" } ], "metrics": { "cvssMetricV31": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H", "baseScore": 6.8, "baseSeverity": "MEDIUM", "attackVector": "PHYSICAL", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "UNCHANGED", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "availabilityImpact": "HIGH" }, "exploitabilityScore": 0.9, "impactScore": 5.9 } ], "cvssMetricV2": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "2.0", "vectorString": "AV:L/AC:L/Au:N/C:P/I:P/A:P", "baseScore": 4.6, "accessVector": "LOCAL", "accessComplexity": "LOW", "authentication": "NONE", "confidentialityImpact": "PARTIAL", "integrityImpact": "PARTIAL", "availabilityImpact": "PARTIAL" }, "baseSeverity": "MEDIUM", "exploitabilityScore": 3.9, "impactScore": 6.4, "acInsufInfo": false, "obtainAllPrivilege": false, "obtainUserPrivilege": false, "obtainOtherPrivilege": false, "userInteractionRequired": false } ] }, "weaknesses": [ { "source": "productcert@siemens.com", "type": "Primary", "description": [ { "lang": "en", "value": "CWE-259" } ] } ], "configurations": [ { "operator": "AND", "nodes": [ { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": true, "criteria": "cpe:2.3:o:siemens:dca_vantage_analyzer_firmware:*:*:*:*:*:*:*:*", "versionEndExcluding": "4.5.0.0", "matchCriteriaId": "2B4F34A3-AB7B-4528-9F27-3073E25A5176" } ] }, { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": false, "criteria": "cpe:2.3:h:siemens:dca_vantage_analyzer:-:*:*:*:*:*:*:*", "matchCriteriaId": "EDE5221F-8629-4DC4-AD8A-93EED671469B" } ] } ] } ], "references": [ { "url": "https://www.siemens-healthineers.com/support-documentation/security-advisory", "source": "productcert@siemens.com", "tags": [ "Vendor Advisory" ] }, { "url": "https://www.siemens-healthineers.com/support-documentation/security-advisory", "source": "af854a3a-2127-422b-91ae-364da2661108", "tags": [ "Vendor Advisory" ] } ] }