{ "id": "CVE-2021-33882", "sourceIdentifier": "cve@mitre.org", "published": "2021-08-25T12:15:13.410", "lastModified": "2021-09-01T15:52:21.417", "vulnStatus": "Analyzed", "descriptions": [ { "lang": "en", "value": "A Missing Authentication for Critical Function vulnerability in B. Braun SpaceCom2 prior to 012U000062 allows a remote attacker to reconfigure the device from an unknown source because of lack of authentication on proprietary networking commands." }, { "lang": "es", "value": "Una vulnerabilidad de Falta de Autenticaci\u00f3n para una Funci\u00f3n Cr\u00edtica en B. Braun SpaceCom2 versiones anteriores a 012U000062, permite a un atacante remoto reconfigurar el dispositivo desde una fuente desconocida debido a una falta de autenticaci\u00f3n en los comandos de red propietarios." } ], "metrics": { "cvssMetricV31": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:N/I:H/A:N", "attackVector": "NETWORK", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "CHANGED", "confidentialityImpact": "NONE", "integrityImpact": "HIGH", "availabilityImpact": "NONE", "baseScore": 8.6, "baseSeverity": "HIGH" }, "exploitabilityScore": 3.9, "impactScore": 4.0 }, { "source": "cve@mitre.org", "type": "Secondary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:C/C:N/I:H/A:N", "attackVector": "NETWORK", "attackComplexity": "HIGH", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "CHANGED", "confidentialityImpact": "NONE", "integrityImpact": "HIGH", "availabilityImpact": "NONE", "baseScore": 6.8, "baseSeverity": "MEDIUM" }, "exploitabilityScore": 2.2, "impactScore": 4.0 } ], "cvssMetricV2": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "2.0", "vectorString": "AV:N/AC:L/Au:N/C:N/I:P/A:N", "accessVector": "NETWORK", "accessComplexity": "LOW", "authentication": "NONE", "confidentialityImpact": "NONE", "integrityImpact": "PARTIAL", "availabilityImpact": "NONE", "baseScore": 5.0 }, "baseSeverity": "MEDIUM", "exploitabilityScore": 10.0, "impactScore": 2.9, "acInsufInfo": false, "obtainAllPrivilege": false, "obtainUserPrivilege": false, "obtainOtherPrivilege": false, "userInteractionRequired": false } ] }, "weaknesses": [ { "source": "nvd@nist.gov", "type": "Primary", "description": [ { "lang": "en", "value": "CWE-306" } ] } ], "configurations": [ { "operator": "AND", "nodes": [ { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": true, "criteria": "cpe:2.3:o:bbraun:spacecom2:*:*:*:*:*:*:*:*", "versionEndExcluding": "012u000062", "matchCriteriaId": "C2D8F2AC-5506-405C-857A-EE060C209A89" } ] }, { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": false, "criteria": "cpe:2.3:h:bbraun:infusomat_large_volume_pump_871305u:-:*:*:*:*:*:*:*", "matchCriteriaId": "A6C9420B-B015-4079-8B60-EF9C9C058193" }, { "vulnerable": false, "criteria": "cpe:2.3:h:bbraun:spacestation_8713142u:-:*:*:*:*:*:*:*", "matchCriteriaId": "D6B64426-0B44-4B68-A80A-2BB6D04B3F88" } ] } ] } ], "references": [ { "url": "https://www.bbraunusa.com/en.htm", "source": "cve@mitre.org", "tags": [ "Broken Link" ] }, { "url": "https://www.mcafee.com/blogs/enterprise/mcafee-enterprise-atr/mcafee-enterprise-atr-uncovers-vulnerabilities-in-globally-used-b-braun-infusion-pump/", "source": "cve@mitre.org", "tags": [ "Third Party Advisory" ] } ] }