{ "id": "CVE-2020-12035", "sourceIdentifier": "ics-cert@hq.dhs.gov", "published": "2020-06-29T14:15:11.397", "lastModified": "2020-07-14T15:34:03.733", "vulnStatus": "Analyzed", "evaluatorComment": "At the time of analysis, information provided did not align with the data (including CVE number) in the advisory.", "descriptions": [ { "lang": "en", "value": "Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The PrismaFlex device contains a hard-coded service password that provides access to biomedical information, device settings, calibration settings, and network configuration. This could allow an attacker to modify device settings and calibration." }, { "lang": "es", "value": "Baxter PrismaFlex todas las versiones, PrisMax todas las versiones anteriores a 3.x, el dispositivo PrismaFlex, contiene una contrase\u00f1a de servicio embebida que proporciona acceso a informaci\u00f3n biom\u00e9dica, configuraci\u00f3n del dispositivo, configuraci\u00f3n de calibraci\u00f3n y configuraci\u00f3n de red. Esto podr\u00eda permitir a un atacante modificar la configuraci\u00f3n y la calibraci\u00f3n del dispositivo" } ], "metrics": { "cvssMetricV31": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:P/AC:L/PR:L/UI:N/S:U/C:N/I:H/A:L", "attackVector": "PHYSICAL", "attackComplexity": "LOW", "privilegesRequired": "LOW", "userInteraction": "NONE", "scope": "UNCHANGED", "confidentialityImpact": "NONE", "integrityImpact": "HIGH", "availabilityImpact": "LOW", "baseScore": 4.9, "baseSeverity": "MEDIUM" }, "exploitabilityScore": 0.7, "impactScore": 4.2 } ], "cvssMetricV2": [ { "source": "nvd@nist.gov", "type": "Primary", "cvssData": { "version": "2.0", "vectorString": "AV:L/AC:L/Au:N/C:N/I:P/A:P", "accessVector": "LOCAL", "accessComplexity": "LOW", "authentication": "NONE", "confidentialityImpact": "NONE", "integrityImpact": "PARTIAL", "availabilityImpact": "PARTIAL", "baseScore": 3.6 }, "baseSeverity": "LOW", "exploitabilityScore": 3.9, "impactScore": 4.9, "acInsufInfo": false, "obtainAllPrivilege": false, "obtainUserPrivilege": false, "obtainOtherPrivilege": false, "userInteractionRequired": false } ] }, "weaknesses": [ { "source": "nvd@nist.gov", "type": "Primary", "description": [ { "lang": "en", "value": "CWE-798" } ] }, { "source": "ics-cert@hq.dhs.gov", "type": "Secondary", "description": [ { "lang": "en", "value": "CWE-287" } ] } ], "configurations": [ { "operator": "AND", "nodes": [ { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": true, "criteria": "cpe:2.3:o:baxter:prismaflex_firmware:*:*:*:*:*:*:*:*", "matchCriteriaId": "B8874804-02E2-4E74-94C6-F103DCDC854E" } ] }, { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": false, "criteria": "cpe:2.3:h:baxter:prismaflex:-:*:*:*:*:*:*:*", "matchCriteriaId": "8E5DB2B2-5D9E-494B-A860-FE2F4BC4C767" } ] } ] }, { "operator": "AND", "nodes": [ { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": true, "criteria": "cpe:2.3:o:baxter:prismax_firmware:*:*:*:*:*:*:*:*", "versionEndExcluding": "3.0", "matchCriteriaId": "8F197DAD-63A3-4598-BBFC-0014707B3A01" } ] }, { "operator": "OR", "negate": false, "cpeMatch": [ { "vulnerable": false, "criteria": "cpe:2.3:h:baxter:prismax:-:*:*:*:*:*:*:*", "matchCriteriaId": "A5EB1A30-A514-437D-B976-451CF003955D" } ] } ] } ], "references": [ { "url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02", "source": "nvd@nist.gov", "tags": [ "Third Party Advisory" ] }, { "url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01", "source": "ics-cert@hq.dhs.gov", "tags": [ "Not Applicable" ] } ] }