{ "id": "CVE-2024-48967", "sourceIdentifier": "productsecurity@baxter.com", "published": "2024-11-14T22:15:17.927", "lastModified": "2024-11-15T13:58:08.913", "vulnStatus": "Awaiting Analysis", "cveTags": [], "descriptions": [ { "lang": "en", "value": "The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance." }, { "lang": "es", "value": "El respirador y la PC de servicio carecen de capacidades de registro de auditor\u00eda suficientes para permitir la detecci\u00f3n de actividad maliciosa y el posterior examen forense. Un atacante con acceso al respirador o a la PC de servicio podr\u00eda, sin ser detectado, realizar cambios no autorizados en la configuraci\u00f3n del respirador que resulten en la divulgaci\u00f3n no autorizada de informaci\u00f3n o tengan impactos no deseados en el rendimiento del dispositivo." } ], "metrics": { "cvssMetricV31": [ { "source": "productsecurity@baxter.com", "type": "Secondary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H", "baseScore": 10.0, "baseSeverity": "CRITICAL", "attackVector": "NETWORK", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "CHANGED", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "availabilityImpact": "HIGH" }, "exploitabilityScore": 3.9, "impactScore": 6.0 } ] }, "weaknesses": [ { "source": "productsecurity@baxter.com", "type": "Secondary", "description": [ { "lang": "en", "value": "CWE-778" } ] } ], "references": [ { "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01", "source": "productsecurity@baxter.com" } ] }