{ "id": "CVE-2024-9832", "sourceIdentifier": "productsecurity@baxter.com", "published": "2024-11-14T21:15:22.593", "lastModified": "2024-11-15T13:58:08.913", "vulnStatus": "Awaiting Analysis", "cveTags": [], "descriptions": [ { "lang": "en", "value": "There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure." }, { "lang": "es", "value": "No existe l\u00edmite en la cantidad de intentos fallidos de inicio de sesi\u00f3n permitidos con la contrase\u00f1a del m\u00e9dico o la contrase\u00f1a del m\u00e9dico con n\u00famero de serie. Un atacante podr\u00eda ejecutar un ataque de fuerza bruta para obtener acceso no autorizado al respirador y luego realizar cambios en la configuraci\u00f3n del dispositivo que podr\u00edan interrumpir el funcionamiento del dispositivo o dar como resultado la divulgaci\u00f3n no autorizada de informaci\u00f3n." } ], "metrics": { "cvssMetricV31": [ { "source": "productsecurity@baxter.com", "type": "Secondary", "cvssData": { "version": "3.1", "vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H", "baseScore": 9.3, "baseSeverity": "CRITICAL", "attackVector": "LOCAL", "attackComplexity": "LOW", "privilegesRequired": "NONE", "userInteraction": "NONE", "scope": "CHANGED", "confidentialityImpact": "HIGH", "integrityImpact": "HIGH", "availabilityImpact": "HIGH" }, "exploitabilityScore": 2.5, "impactScore": 6.0 } ] }, "weaknesses": [ { "source": "productsecurity@baxter.com", "type": "Secondary", "description": [ { "lang": "en", "value": "CWE-307" } ] } ], "references": [ { "url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01", "source": "productsecurity@baxter.com" } ] }