"lang": "en",
"value": "The CMS800 device fails while attempting to parse malformed network data sent by a threat actor. A threat actor with network access can remotely issue a specially formatted UDP request that will cause the entire device to crash and require a physical reboot. A UDP broadcast request could be sent that causes a mass denial-of-service attack on all CME8000 devices connected to the same network."
},
{
"lang": "es",
"value": "El dispositivo CMS800 falla mientras intenta analizar los datos de red malformados enviados por un actor de la amenaza. Un actor de la amenaza con acceso a la red puede emitir de forma remota una petici\u00f3n UDP con un formato especial que causar\u00e1 el bloqueo de todo el dispositivo y requerir\u00e1 un reinicio f\u00edsico. Puede enviarse una petici\u00f3n de difusi\u00f3n UDP que cause un ataque de denegaci\u00f3n de servicio masivo en todos los dispositivos CME8000 conectados a la misma red"
}
],
"metrics": {
"cvssMetricV31": [
{
"source": "nvd@nist.gov",
"type": "Primary",
"cvssData": {
"version": "3.1",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"attackVector": "NETWORK",
"attackComplexity": "LOW",
"privilegesRequired": "NONE",
"userInteraction": "NONE",
"scope": "UNCHANGED",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH"
},
"exploitabilityScore": 3.9,
"impactScore": 3.6
},
{
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary",
"cvssData": {
"version": "3.1",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:N/I:N/A:H",
"attackVector": "NETWORK",
"attackComplexity": "LOW",
"privilegesRequired": "NONE",
"userInteraction": "NONE",
"scope": "UNCHANGED",
"confidentialityImpact": "NONE",
"integrityImpact": "NONE",
"availabilityImpact": "HIGH",
"baseScore": 7.5,
"baseSeverity": "HIGH"
},
"exploitabilityScore": 3.9,
"impactScore": 3.6
}
]
},
"weaknesses": [
{
"source": "nvd@nist.gov",
"type": "Primary",
"description": [
{
"lang": "en",
"value": "NVD-CWE-Other"
}
]
},
{
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary",
"description": [
{
"lang": "en",
"value": "CWE-400"
}
]
}
],
"configurations": [
{
"operator": "AND",
"nodes": [
{
"operator": "OR",
"negate": false,
"cpeMatch": [
{
"vulnerable": true,
"criteria": "cpe:2.3:o:contechealth:cms8000_firmware:-:*:*:*:*:*:*:*",
"matchCriteriaId": "2C197D62-6F35-4B87-A721-BDB696EA240F"
}
]
},
{
"operator": "OR",
"negate": false,
"cpeMatch": [
{
"vulnerable": false,
"criteria": "cpe:2.3:h:contechealth:cms8000:-:*:*:*:*:*:*:*",
"matchCriteriaId": "3A0CD9FA-68D7-4EEE-93A5-97275D84E2D3"
}
]
}
]
}
],