"id": "CVE-2025-0626",
|
|
"sourceIdentifier": "ics-cert@hq.dhs.gov",
|
|
"published": "2025-01-30T19:15:14.010",
|
|
"lastModified": "2025-03-01T18:15:34.140",
|
|
"vulnStatus": "Awaiting Analysis",
|
|
"cveTags": [],
|
|
"descriptions": [
|
|
{
|
|
"lang": "en",
|
|
"value": "The \"monitor\" binary in the firmware of the affected product attempts to mount to a hard-coded, routable IP address, bypassing existing device network settings to do so. The function also enables the network interface of the device if it is disabled. The function is triggered by attempting to update the device from the user menu. This could serve as a backdoor to the device, and could lead to a malicious actor being able to upload and overwrite files on the device."
|
|
},
|
|
{
|
|
"lang": "es",
|
|
"value": "El producto afectado env\u00eda solicitudes de acceso remoto a una direcci\u00f3n IP codificada, omitiendo la configuraci\u00f3n de red existente del dispositivo para hacerlo. Esto podr\u00eda funcionar como una puerta trasera y permitir que un actor malintencionado pueda cargar y sobrescribir archivos en el dispositivo."
|
|
}
|
|
],
|
|
"metrics": {
|
|
"cvssMetricV40": [
|
|
{
|
|
"source": "ics-cert@hq.dhs.gov",
|
|
"type": "Secondary",
|
|
"cvssData": {
|
|
"version": "4.0",
|
|
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:P/VC:H/VI:H/VA:H/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X",
|
|
"baseScore": 7.7,
|
|
"baseSeverity": "HIGH",
|
|
"attackVector": "NETWORK",
|
|
"attackComplexity": "LOW",
|
|
"attackRequirements": "PRESENT",
|
|
"privilegesRequired": "NONE",
|
|
"userInteraction": "PASSIVE",
|
|
"vulnConfidentialityImpact": "HIGH",
|
|
"vulnIntegrityImpact": "HIGH",
|
|
"vulnAvailabilityImpact": "HIGH",
|
|
"subConfidentialityImpact": "NONE",
|
|
"subIntegrityImpact": "NONE",
|
|
"subAvailabilityImpact": "NONE",
|
|
"exploitMaturity": "NOT_DEFINED",
|
|
"confidentialityRequirement": "NOT_DEFINED",
|
|
"integrityRequirement": "NOT_DEFINED",
|
|
"availabilityRequirement": "NOT_DEFINED",
|
|
"modifiedAttackVector": "NOT_DEFINED",
|
|
"modifiedAttackComplexity": "NOT_DEFINED",
|
|
"modifiedAttackRequirements": "NOT_DEFINED",
|
|
"modifiedPrivilegesRequired": "NOT_DEFINED",
|
|
"modifiedUserInteraction": "NOT_DEFINED",
|
|
"modifiedVulnConfidentialityImpact": "NOT_DEFINED",
|
|
"modifiedVulnIntegrityImpact": "NOT_DEFINED",
|
|
"modifiedVulnAvailabilityImpact": "NOT_DEFINED",
|
|
"modifiedSubConfidentialityImpact": "NOT_DEFINED",
|
|
"modifiedSubIntegrityImpact": "NOT_DEFINED",
|
|
"modifiedSubAvailabilityImpact": "NOT_DEFINED",
|
|
"Safety": "NOT_DEFINED",
|
|
"Automatable": "NOT_DEFINED",
|
|
"Recovery": "NOT_DEFINED",
|
|
"valueDensity": "NOT_DEFINED",
|
|
"vulnerabilityResponseEffort": "NOT_DEFINED",
|
|
"providerUrgency": "NOT_DEFINED"
|
|
}
|
|
}
|
|
],
|
|
"cvssMetricV31": [
|
|
{
|
|
"source": "ics-cert@hq.dhs.gov",
|
|
"type": "Secondary",
|
|
"cvssData": {
|
|
"version": "3.1",
|
|
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:R/S:U/C:H/I:H/A:H",
|
|
"baseScore": 7.5,
|
|
"baseSeverity": "HIGH",
|
|
"attackVector": "NETWORK",
|
|
"attackComplexity": "HIGH",
|
|
"privilegesRequired": "NONE",
|
|
"userInteraction": "REQUIRED",
|
|
"scope": "UNCHANGED",
|
|
"confidentialityImpact": "HIGH",
|
|
"integrityImpact": "HIGH",
|
|
"availabilityImpact": "HIGH"
|
|
},
|
|
"exploitabilityScore": 1.6,
|
|
"impactScore": 5.9
|
|
}
|
|
]
|
|
},
|
|
"weaknesses": [
|
|
{
|
|
"source": "ics-cert@hq.dhs.gov",
|
|
"type": "Secondary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-912"
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"references": [
|
|
{
|
|
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01",
|
|
"source": "ics-cert@hq.dhs.gov"
|
|
},
|
|
{
|
|
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication",
|
|
"source": "ics-cert@hq.dhs.gov"
|
|
},
|
|
{
|
|
"url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/",
|
|
"source": "af854a3a-2127-422b-91ae-364da2661108"
|
|
},
|
|
{
|
|
"url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor",
|
|
"source": "af854a3a-2127-422b-91ae-364da2661108"
|
|
},
|
|
{
|
|
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication",
|
|
"source": "af854a3a-2127-422b-91ae-364da2661108"
|
|
}
|
|
]
|
|
} |