"id": "CVE-2025-0683",
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"published": "2025-01-30T19:15:14.460",
"lastModified": "2025-01-31T18:15:37.750",
"vulnStatus": "Awaiting Analysis",
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "In its default configuration, Contec Health CMS8000 Patient Monitor transmits plain-text \npatient data to a hard-coded public IP address when a patient is hooked \nup to the monitor. This could lead to a leakage of confidential patient \ndata to any device with that IP address or an attacker in a \nmachine-in-the-middle scenario."
},
{
"lang": "es",
"value": "En su configuraci\u00f3n predeterminada, el producto afectado transmite datos del paciente en texto plano a una direcci\u00f3n IP p\u00fablica codificada cuando un paciente est\u00e1 conectado al monitor. Esto podr\u00eda provocar una fuga de datos confidenciales del paciente a cualquier dispositivo con esa direcci\u00f3n IP o a un atacante en un escenario de m\u00e1quina intermedia."
}
],
"metrics": {
"cvssMetricV40": [
{
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary",
"cvssData": {
"version": "4.0",
"vectorString": "CVSS:4.0/AV:N/AC:L/AT:P/PR:N/UI:N/VC:H/VI:N/VA:N/SC:N/SI:N/SA:N/E:X/CR:X/IR:X/AR:X/MAV:X/MAC:X/MAT:X/MPR:X/MUI:X/MVC:X/MVI:X/MVA:X/MSC:X/MSI:X/MSA:X/S:X/AU:X/R:X/V:X/RE:X/U:X",
"baseScore": 8.2,
"baseSeverity": "HIGH",
"attackVector": "NETWORK",
"attackComplexity": "LOW",
"attackRequirements": "PRESENT",
"privilegesRequired": "NONE",
"userInteraction": "NONE",
"vulnConfidentialityImpact": "HIGH",
"vulnIntegrityImpact": "NONE",
"vulnAvailabilityImpact": "NONE",
"subConfidentialityImpact": "NONE",
"subIntegrityImpact": "NONE",
"subAvailabilityImpact": "NONE",
"exploitMaturity": "NOT_DEFINED",
"confidentialityRequirement": "NOT_DEFINED",
"integrityRequirement": "NOT_DEFINED",
"availabilityRequirement": "NOT_DEFINED",
"modifiedAttackVector": "NOT_DEFINED",
"modifiedAttackComplexity": "NOT_DEFINED",
"modifiedAttackRequirements": "NOT_DEFINED",
"modifiedPrivilegesRequired": "NOT_DEFINED",
"modifiedUserInteraction": "NOT_DEFINED",
"modifiedVulnConfidentialityImpact": "NOT_DEFINED",
"modifiedVulnIntegrityImpact": "NOT_DEFINED",
"modifiedVulnAvailabilityImpact": "NOT_DEFINED",
"modifiedSubConfidentialityImpact": "NOT_DEFINED",
"modifiedSubIntegrityImpact": "NOT_DEFINED",
"modifiedSubAvailabilityImpact": "NOT_DEFINED",
"Safety": "NOT_DEFINED",
"Automatable": "NOT_DEFINED",
"Recovery": "NOT_DEFINED",
"valueDensity": "NOT_DEFINED",
"vulnerabilityResponseEffort": "NOT_DEFINED",
"providerUrgency": "NOT_DEFINED"
}
}
],
"cvssMetricV31": [
{
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary",
"cvssData": {
"version": "3.1",
"vectorString": "CVSS:3.1/AV:N/AC:H/PR:N/UI:N/S:U/C:H/I:N/A:N",
"baseScore": 5.9,
"baseSeverity": "MEDIUM",
"attackVector": "NETWORK",
"attackComplexity": "HIGH",
"privilegesRequired": "NONE",
"userInteraction": "NONE",
"scope": "UNCHANGED",
"confidentialityImpact": "HIGH",
"integrityImpact": "NONE",
"availabilityImpact": "NONE"
},
"exploitabilityScore": 2.2,
"impactScore": 3.6
}
]
},
"weaknesses": [
{
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary",
"description": [
{
"lang": "en",
"value": "CWE-359"
}
]
}
],
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-25-030-01",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication",
"source": "ics-cert@hq.dhs.gov"
},
{
"url": "https://www.bleepingcomputer.com/news/security/backdoor-found-in-two-healthcare-patient-monitors-linked-to-ip-in-china/",
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://www.cisa.gov/resources-tools/resources/contec-cms8000-contains-backdoor",
"source": "af854a3a-2127-422b-91ae-364da2661108"
},
{
"url": "https://www.fda.gov/medical-devices/safety-communications/cybersecurity-vulnerabilities-certain-patient-monitors-contec-and-epsimed-fda-safety-communication",
"source": "af854a3a-2127-422b-91ae-364da2661108"
}
]
}