"lang": "en",
|
|
"value": "Hospira Plum A+ Infusion System version 13.4 and prior, Plum A+3 Infusion System version 13.6 and prior, and Symbiq Infusion System, version 3.13 and prior give unauthenticated users root privileges on Port 23/TELNET by default. An unauthorized user could issue commands to the pump. Hospira recommends that customers close Port 23/TELNET on the affected devices. Hospira has also released the Plum 360 Infusion System which is not vulnerable to this issue."
|
|
},
|
|
{
|
|
"lang": "es",
|
|
"value": "Hospira Plum A+ Infusion System versi\u00f3n 13.4 y anteriores, Plum A+3 Infusion System en su versi\u00f3n 13.6 y anteriores y Symbiq Infusion System, en su versi\u00f3n 3.13 y anteriores proporcionan privilegios root a usuarios no autenticados en el Puerto 23/TELNET por defecto. Un usuario no autorizado podr\u00eda enviar comandos a la bomba. Hospira recomienda con los usuarios cierren el Puerto 23/TELNET en los dispositivos afectados. Hospira ha lanzado tambi\u00e9n el Plum 360 Infusion System que no es vulnerable a este problema."
|
|
}
|
|
],
|
|
"metrics": {
|
|
"cvssMetricV30": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "3.0",
|
|
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
|
|
"attackVector": "NETWORK",
|
|
"attackComplexity": "LOW",
|
|
"privilegesRequired": "NONE",
|
|
"userInteraction": "NONE",
|
|
"scope": "UNCHANGED",
|
|
"confidentialityImpact": "HIGH",
|
|
"integrityImpact": "HIGH",
|
|
"availabilityImpact": "HIGH",
|
|
"baseScore": 9.8,
|
|
"baseSeverity": "CRITICAL"
|
|
},
|
|
"exploitabilityScore": 3.9,
|
|
"impactScore": 5.9
|
|
}
|
|
],
|
|
"cvssMetricV2": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "2.0",
|
|
"vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
|
|
"accessVector": "NETWORK",
|
|
"accessComplexity": "LOW",
|
|
"authentication": "NONE",
|
|
"confidentialityImpact": "COMPLETE",
|
|
"integrityImpact": "COMPLETE",
|
|
"availabilityImpact": "COMPLETE",
|
|
"baseScore": 10.0
|
|
},
|
|
"baseSeverity": "HIGH",
|
|
"exploitabilityScore": 10.0,
|
|
"impactScore": 10.0,
|
|
"acInsufInfo": false,
|
|
"obtainAllPrivilege": false,
|
|
"obtainUserPrivilege": false,
|
|
"obtainOtherPrivilege": false,
|
|
"userInteractionRequired": false
|
|
}
|
|
]
|
|
},
|
|
"weaknesses": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-285"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"source": "ics-cert@hq.dhs.gov",
|
|
"type": "Secondary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-285"
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"configurations": [
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:pifzer:plum_a\\+_infusion_system_firmware:*:*:*:*:*:*:*:*",
|
|
"versionEndIncluding": "13.4",
|
|
"matchCriteriaId": "61C1A9DD-F143-4D0C-871C-B6CD7AF9DAB2"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:pifzer:plum_a\\+_infusion_system:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "1DB2490B-0318-4770-BF45-CD7527F15D7F"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:pifzer:plum_a\\+3_infusion_system_firmware:*:*:*:*:*:*:*:*",
|
|
"versionEndIncluding": "13.6",
|
|
"matchCriteriaId": "8536E705-89E8-47CB-9567-6AD65FBA0F1B"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:pifzer:plum_a\\+3_infusion_system:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "423AA561-8E38-4378-814B-1008B96F27A6"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:pifzer:symbiq_infusion_system_firmware:*:*:*:*:*:*:*:*",
|
|
"versionEndIncluding": "3.13",
|
|
"matchCriteriaId": "C6DD5141-72AB-4694-8595-B4BED8EC7773"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:pifzer:symbiq_infusion_system:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "5C295E1A-BF60-476D-B972-5C5C28D7633B"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"references": [
|
|
{
|
|
"url": "https://ics-cert.us-cert.gov/advisories/ICSA-15-161-01",
|
|
"source": "ics-cert@hq.dhs.gov",
|
|
"tags": [
|
|
"Mitigation",
|
|
"US Government Resource",
|
|
"Third Party Advisory"
|
|
]
|
|
}
|
|
]
|
|
} |