"id": "CVE-2018-4846",
|
|
"sourceIdentifier": "productcert@siemens.com",
|
|
"published": "2018-06-26T18:29:00.870",
|
|
"lastModified": "2019-10-09T23:41:01.890",
|
|
"vulnStatus": "Modified",
|
|
"descriptions": [
|
|
{
|
|
"lang": "en",
|
|
"value": "A vulnerability has been identified in RAPIDLab 1200 systems / RAPIDPoint 400 systems / RAPIDPoint 500 systems (All versions_without_ use of Siemens Healthineers Informatics products), RAPIDLab 1200 Series (All versions < V3.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions >= V3.0 _with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (V2.4.X_with_ Siemens Healthineers Informatics products), RAPIDPoint 500 systems (All versions =< V2.3 _with_ Siemens Healthineers Informatics products), RAPIDPoint 400 systems (All versions _with_ Siemens Healthineers Informatics products). A factory account with hardcoded password might allow attackers access to the device over port 5900/tcp. Successful exploitation requires no user interaction or privileges and impacts the confidentiality, integrity, and availability of the affected device. At the time of advisory publication, no public exploitation of this security vulnerability is known. Siemens Healthineers confirms the security vulnerability and provides mitigations to resolve the security issue."
|
|
},
|
|
{
|
|
"lang": "es",
|
|
"value": "Se ha identificado una vulnerabilidad en los sistemas 1200 de RAPIDLab, los sistemas 400 de RAPIDPoint, los sistemas 500 de RAPIDPoint (todas las versiones sin usar productos Siemens Healthineers Informatics), las series 1200 de RAPIDLab (todas las versiones anteriores a la V3.3 con productos Siemens Healthineers Informatics), los sistemas 500 de RAPIDPoint (versiones 3.0 y posteriores con productos Siemens Healthineers Informatics), los sistemas 500 de RAPIDPoint (versiones 3.0 y posteriores con productos Siemens Healthineers Informatics), los sistemas 500 de RAPIDPoint (versiones 2.4.X con productos Siemens Healthineers Informatics) y los sistemas 400 de RAPIDPoint (todas las versiones con productos Siemens Healthineers Informatics). Una cuenta de f\u00e1brica con contrase\u00f1a embebida podr\u00eda permitir que los atacantes accedan al dispositivo por el puerto 5900/tcp. La explotaci\u00f3n con \u00e9xito no requiere de interacci\u00f3n o privilegios de usuario e impacta en la confidencialidad, integridad y disponibilidad del dispositivo afectado. En el momento de la publicaci\u00f3n del advisory, no se conoce ninguna explotaci\u00f3n p\u00fablica de la vulnerabilidad de seguridad. Siemens Healthineers confirma la vulnerabilidad de seguridad y proporciona mitigaciones para resolver el problema de seguridad."
|
|
}
|
|
],
|
|
"metrics": {
|
|
"cvssMetricV30": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "3.0",
|
|
"vectorString": "CVSS:3.0/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
|
|
"attackVector": "NETWORK",
|
|
"attackComplexity": "LOW",
|
|
"privilegesRequired": "NONE",
|
|
"userInteraction": "NONE",
|
|
"scope": "UNCHANGED",
|
|
"confidentialityImpact": "HIGH",
|
|
"integrityImpact": "HIGH",
|
|
"availabilityImpact": "HIGH",
|
|
"baseScore": 9.8,
|
|
"baseSeverity": "CRITICAL"
|
|
},
|
|
"exploitabilityScore": 3.9,
|
|
"impactScore": 5.9
|
|
}
|
|
],
|
|
"cvssMetricV2": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "2.0",
|
|
"vectorString": "AV:N/AC:L/Au:N/C:C/I:C/A:C",
|
|
"accessVector": "NETWORK",
|
|
"accessComplexity": "LOW",
|
|
"authentication": "NONE",
|
|
"confidentialityImpact": "COMPLETE",
|
|
"integrityImpact": "COMPLETE",
|
|
"availabilityImpact": "COMPLETE",
|
|
"baseScore": 10.0
|
|
},
|
|
"baseSeverity": "HIGH",
|
|
"exploitabilityScore": 10.0,
|
|
"impactScore": 10.0,
|
|
"acInsufInfo": false,
|
|
"obtainAllPrivilege": false,
|
|
"obtainUserPrivilege": false,
|
|
"obtainOtherPrivilege": false,
|
|
"userInteractionRequired": false
|
|
}
|
|
]
|
|
},
|
|
"weaknesses": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-798"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"source": "productcert@siemens.com",
|
|
"type": "Secondary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-798"
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"configurations": [
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:siemens:rapidpoint_400_firmware:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "D3BD8C53-0B6B-4961-AF6E-302923B34118"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:siemens:rapidpoint_400:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "6081ED98-FD21-4C35-9E6F-95089EDB2905"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:siemens:rapidpoint_500_firmware:*:*:*:*:*:*:*:*",
|
|
"versionEndIncluding": "2.3",
|
|
"matchCriteriaId": "15D5C974-4C15-450B-B9EF-51AE84A1741A"
|
|
},
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:siemens:rapidpoint_500_firmware:*:*:*:*:*:*:*:*",
|
|
"versionStartIncluding": "3.0",
|
|
"matchCriteriaId": "B2624D5C-08DF-4D18-8EC5-51509C2BFCE9"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:siemens:rapidpoint_500:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "9A4806F5-0292-45BA-96B9-422958D0CB0F"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:siemens:rapidlab_1200_firmware:*:*:*:*:*:*:*:*",
|
|
"versionEndExcluding": "3.3",
|
|
"matchCriteriaId": "8A093C69-118A-44F9-98ED-B2B103B1EFBB"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:siemens:rapidlab_1200:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "757CBCB7-BE66-43E2-BD7D-55B67A649EBC"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"references": [
|
|
{
|
|
"url": "https://cert-portal.siemens.com/productcert/pdf/ssa-755010.pdf",
|
|
"source": "productcert@siemens.com",
|
|
"tags": [
|
|
"Third Party Advisory"
|
|
]
|
|
}
|
|
]
|
|
} |