"id": "CVE-2020-12037",
|
|
"sourceIdentifier": "ics-cert@hq.dhs.gov",
|
|
"published": "2020-06-29T14:15:11.537",
|
|
"lastModified": "2020-07-14T15:41:13.167",
|
|
"vulnStatus": "Analyzed",
|
|
"evaluatorComment": "At the time of analysis, information provided did not align with the data (including CVE number) in the advisory.",
|
|
"descriptions": [
|
|
{
|
|
"lang": "en",
|
|
"value": "Baxter PrismaFlex all versions, PrisMax all versions prior to 3.x, The affected devices do not implement data-in-transit encryption (e.g., TLS/SSL) when configured to send treatment data to a PDMS (Patient Data Management System) or an EMR (Electronic Medical Record) system. An attacker could observe sensitive data sent from the device."
|
|
},
|
|
{
|
|
"lang": "es",
|
|
"value": "Baxter PrismaFlex todas las versiones, PrisMax todas las versiones anteriores a 3.x, los dispositivos afectados no implementan cifrado de datos en tr\u00e1nsito (por ejemplo, TLS/SSL) cuando se configuran para enviar datos de tratamiento a un PDMS (Patient Data Management System) o un sistema EMR (Electronic Medical Record). Un atacante podr\u00eda observar datos confidenciales enviados desde el dispositivo"
|
|
}
|
|
],
|
|
"metrics": {
|
|
"cvssMetricV31": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "3.1",
|
|
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
|
|
"attackVector": "NETWORK",
|
|
"attackComplexity": "LOW",
|
|
"privilegesRequired": "NONE",
|
|
"userInteraction": "NONE",
|
|
"scope": "UNCHANGED",
|
|
"confidentialityImpact": "HIGH",
|
|
"integrityImpact": "NONE",
|
|
"availabilityImpact": "NONE",
|
|
"baseScore": 7.5,
|
|
"baseSeverity": "HIGH"
|
|
},
|
|
"exploitabilityScore": 3.9,
|
|
"impactScore": 3.6
|
|
}
|
|
],
|
|
"cvssMetricV2": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "2.0",
|
|
"vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N",
|
|
"accessVector": "NETWORK",
|
|
"accessComplexity": "LOW",
|
|
"authentication": "NONE",
|
|
"confidentialityImpact": "PARTIAL",
|
|
"integrityImpact": "NONE",
|
|
"availabilityImpact": "NONE",
|
|
"baseScore": 5.0
|
|
},
|
|
"baseSeverity": "MEDIUM",
|
|
"exploitabilityScore": 10.0,
|
|
"impactScore": 2.9,
|
|
"acInsufInfo": false,
|
|
"obtainAllPrivilege": false,
|
|
"obtainUserPrivilege": false,
|
|
"obtainOtherPrivilege": false,
|
|
"userInteractionRequired": false
|
|
}
|
|
]
|
|
},
|
|
"weaknesses": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-319"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"source": "ics-cert@hq.dhs.gov",
|
|
"type": "Secondary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-259"
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"configurations": [
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:baxter:prismaflex_firmware:*:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "B8874804-02E2-4E74-94C6-F103DCDC854E"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:baxter:prismaflex:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "8E5DB2B2-5D9E-494B-A860-FE2F4BC4C767"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:baxter:prismax_firmware:*:*:*:*:*:*:*:*",
|
|
"versionEndExcluding": "3.0",
|
|
"matchCriteriaId": "8F197DAD-63A3-4598-BBFC-0014707B3A01"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:baxter:prismax:-:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "A5EB1A30-A514-437D-B976-451CF003955D"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"references": [
|
|
{
|
|
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-20-170-02",
|
|
"source": "nvd@nist.gov",
|
|
"tags": [
|
|
"Third Party Advisory"
|
|
]
|
|
},
|
|
{
|
|
"url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-01",
|
|
"source": "ics-cert@hq.dhs.gov",
|
|
"tags": [
|
|
"Not Applicable"
|
|
]
|
|
}
|
|
]
|
|
} |