"id": "CVE-2020-12040",
"sourceIdentifier": "ics-cert@hq.dhs.gov",
"published": "2020-06-29T14:15:11.677",
"lastModified": "2020-07-09T17:58:57.613",
"vulnStatus": "Analyzed",
"descriptions": [
{
"lang": "en",
"value": "Sigma Spectrum Infusion System v's6.x (model 35700BAX) and Baxter Spectrum Infusion System Version(s) 8.x (model 35700BAX2) at the application layer uses an unauthenticated clear-text communication channel to send and receive system status and operational data. This could allow an attacker that has circumvented network security measures to view sensitive non-private data or to perform a man-in-the-middle attack."
},
{
"lang": "es",
"value": "Sigma Spectrum Infusion System versiones v's6.x (modelo 35700BAX) y Baxter Spectrum Infusion System versiones 8.x (modelo 35700BAX2), en la capa de aplicaci\u00f3n utilizan un canal de comunicaci\u00f3n de texto sin cifrar no autenticado para enviar y recibir el estado del sistema y los datos operativos. Esto podr\u00eda permitir a un atacante que haya eludido las medidas de seguridad de la red visualizar datos confidenciales no privados o llevar a cabo un ataque de tipo man-in-the-middle"
}
],
"metrics": {
"cvssMetricV31": [
{
"source": "nvd@nist.gov",
"type": "Primary",
"cvssData": {
"version": "3.1",
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:U/C:H/I:H/A:H",
"attackVector": "NETWORK",
"attackComplexity": "LOW",
"privilegesRequired": "NONE",
"userInteraction": "NONE",
"scope": "UNCHANGED",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"availabilityImpact": "HIGH",
"baseScore": 9.8,
"baseSeverity": "CRITICAL"
},
"exploitabilityScore": 3.9,
"impactScore": 5.9
}
],
"cvssMetricV2": [
{
"source": "nvd@nist.gov",
"type": "Primary",
"cvssData": {
"version": "2.0",
"vectorString": "AV:N/AC:L/Au:N/C:P/I:N/A:N",
"accessVector": "NETWORK",
"accessComplexity": "LOW",
"authentication": "NONE",
"confidentialityImpact": "PARTIAL",
"integrityImpact": "NONE",
"availabilityImpact": "NONE",
"baseScore": 5.0
},
"baseSeverity": "MEDIUM",
"exploitabilityScore": 10.0,
"impactScore": 2.9,
"acInsufInfo": false,
"obtainAllPrivilege": false,
"obtainUserPrivilege": false,
"obtainOtherPrivilege": false,
"userInteractionRequired": false
}
]
},
"weaknesses": [
{
"source": "nvd@nist.gov",
"type": "Primary",
"description": [
{
"lang": "en",
"value": "CWE-319"
}
]
},
{
"source": "ics-cert@hq.dhs.gov",
"type": "Secondary",
"description": [
{
"lang": "en",
"value": "CWE-319"
}
]
}
],
"configurations": [
{
"operator": "AND",
"nodes": [
{
"operator": "OR",
"negate": false,
"cpeMatch": [
{
"vulnerable": true,
"criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:*:*:*:*:*:*:*:*",
"versionStartIncluding": "6.0",
"versionEndIncluding": "6.05",
"matchCriteriaId": "0FCC49AB-B0AD-44FF-ACEB-3C26EA4F2C3C"
},
{
"vulnerable": true,
"criteria": "cpe:2.3:o:baxter:sigma_spectrum_infusion_system_firmware:8.0:*:*:*:*:*:*:*",
"matchCriteriaId": "72D0F7CB-3D24-4A8D-826D-ACB20ACBEB1C"
}
]
},
{
"operator": "OR",
"negate": false,
"cpeMatch": [
{
"vulnerable": false,
"criteria": "cpe:2.3:h:baxter:sigma_spectrum_infusion_system:-:*:*:*:*:*:*:*",
"matchCriteriaId": "49E25260-EC14-4E98-A86B-CBBE47E26AE5"
}
]
}
]
}
],
"references": [
{
"url": "https://www.us-cert.gov/ics/advisories/icsma-20-170-04",
"source": "ics-cert@hq.dhs.gov",
"tags": [
"Third Party Advisory",
"US Government Resource"
]
}
]
}