2024-12-08 03:06:42 +00:00

60 lines
2.0 KiB
JSON

{
"id": "CVE-2024-48970",
"sourceIdentifier": "productsecurity@baxter.com",
"published": "2024-11-14T22:15:18.137",
"lastModified": "2024-11-15T13:58:08.913",
"vulnStatus": "Awaiting Analysis",
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "The ventilator's microcontroller lacks memory protection. An attacker could connect to the internal JTAG interface and read or write to flash memory using an off-the-shelf debugging tool, which could disrupt the function of the device and/or cause unauthorized information disclosure."
},
{
"lang": "es",
"value": "El microcontrolador del respirador carece de protecci\u00f3n de memoria. Un atacante podr\u00eda conectarse a la interfaz JTAG interna y leer o escribir en la memoria flash mediante una herramienta de depuraci\u00f3n comercial, lo que podr\u00eda interrumpir el funcionamiento del dispositivo o provocar la divulgaci\u00f3n no autorizada de informaci\u00f3n."
}
],
"metrics": {
"cvssMetricV31": [
{
"source": "productsecurity@baxter.com",
"type": "Secondary",
"cvssData": {
"version": "3.1",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"attackVector": "LOCAL",
"attackComplexity": "LOW",
"privilegesRequired": "NONE",
"userInteraction": "NONE",
"scope": "CHANGED",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"availabilityImpact": "HIGH"
},
"exploitabilityScore": 2.5,
"impactScore": 6.0
}
]
},
"weaknesses": [
{
"source": "productsecurity@baxter.com",
"type": "Secondary",
"description": [
{
"lang": "en",
"value": "CWE-1191"
}
]
}
],
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01",
"source": "productsecurity@baxter.com"
}
]
}