"CVE_data_meta": {
"ASSIGNER": "ics-cert@hq.dhs.gov",
"DATE_PUBLIC": "2021-09-30T21:02:00.000Z",
"ID": "CVE-2021-38394",
"STATE": "PUBLIC",
"TITLE": "Missing Protection against Hardware Reverse Engineering Using Integrated Circuit Imaging Techniques for Boston Scientific Zoom Latitude"
},
"affects": {
"vendor": {
"vendor_data": [
{
"product": {
"product_data": [
{
"product_name": "ZOOM LATITUDE",
"version": {
"version_data": [
{
"version_value": "Model 3120"
}
]
}
}
]
},
"vendor_name": "Boston Scientific"
}
]
}
},
"credit": [
{
"lang": "eng",
"value": "Endres Puschner - Max Planck Institute for Security and Privacy, Bochum, Christoph Saatjohann - FH M\u00fcnster University of Applied Sciences, Christian Dresen - FH M\u00fcnster University of Applied Sciences, and Markus Willing - University of Muenster, discovered these issues as part of broader academic research of cardiac devices and reported them to Boston Scientific."
}
],
"data_format": "MITRE",
"data_type": "CVE",
"data_version": "4.0",
"description": {
"description_data": [
{
"lang": "eng",
"value": "An attacker with physical access to the device can extract the binary that checks for the hardware key and reverse engineer it, which could be used to create a physical duplicate of a valid hardware key. The hardware key allows access to special settings when inserted."
}
]
},
"generator": {
"engine": "Vulnogram 0.0.9"
},
"impact": {
"cvss": {
"attackComplexity": "HIGH",
"attackVector": "PHYSICAL",
"availabilityImpact": "LOW",
"baseScore": 6.2,
"baseSeverity": "MEDIUM",
"confidentialityImpact": "LOW",
"integrityImpact": "HIGH",
"privilegesRequired": "NONE",
"scope": "CHANGED",
"userInteraction": "REQUIRED",
"vectorString": "CVSS:3.1/AV:P/AC:H/PR:N/UI:R/S:C/C:L/I:H/A:L",
"version": "3.1"
}
},
"problemtype": {
"problemtype_data": [
{
"description": [
{
"lang": "eng",
"value": "CWE-1278 "
}
]
}
]
},
"references": {
"reference_data": [
{
"refsource": "MISC",
"url": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01",
"name": "https://us-cert.cisa.gov/ics/advisories/icsma-21-273-01"
}
]
},
"source": {
"advisory": "ICSMA-21-273-01",
"defect": [
"CWE-1278"
],
"discovery": "EXTERNAL"
},
"work_around": [
{
"lang": "eng",
"value": "Boston Scientific is in the process of transitioning all users to a replacement programmer with enhanced security, the LATITUDE Programming System, Model 3300. Boston Scientific will not issue a product update to address the identified vulnerabilities in the ZOOM LATITUDE Programming System, Model 3120."
}
]
}