mirror of
https://github.com/fkie-cad/nvd-json-data-feeds.git
synced 2025-05-28 01:02:25 +00:00
116 lines
3.9 KiB
JSON
116 lines
3.9 KiB
JSON
{
|
|
"id": "CVE-2020-15486",
|
|
"sourceIdentifier": "cve@mitre.org",
|
|
"published": "2020-08-26T16:15:12.683",
|
|
"lastModified": "2021-07-21T11:39:23.747",
|
|
"vulnStatus": "Analyzed",
|
|
"descriptions": [
|
|
{
|
|
"lang": "en",
|
|
"value": "An issue was discovered on Dr Trust ECG Pen 2.00.08 devices. Because the Bluetooth LE support is implemented without a requirement for pairing or security, any attacker can access the GATT server of the device and can sniff the data being broadcasted while a measurement is being done. Also, saved data can also be extracted over a Bluetooth connection. In addition, an attacker can launch a man-in-the-middle attack against data integrity."
|
|
},
|
|
{
|
|
"lang": "es",
|
|
"value": "Se detect\u00f3 un problema en los dispositivos Dr Trust ECG Pen versi\u00f3n 2.00.08. Porque la compatibilidad con Bluetooth LE es implementada sin un requisito de emparejamiento o seguridad, cualquier atacante puede acceder al servidor GATT del dispositivo y puede rastrear los datos que han sido transmitidos mientras ha sido realizada una medici\u00f3n. Adem\u00e1s, los datos guardados tambi\u00e9n pueden ser extra\u00eddos a trav\u00e9s de una conexi\u00f3n Bluetooth. Adem\u00e1s, un atacante puede iniciar un ataque de tipo man-in-the-middle contra la integridad de los datos"
|
|
}
|
|
],
|
|
"metrics": {
|
|
"cvssMetricV31": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "3.1",
|
|
"vectorString": "CVSS:3.1/AV:A/AC:L/PR:N/UI:N/S:U/C:H/I:N/A:N",
|
|
"attackVector": "ADJACENT_NETWORK",
|
|
"attackComplexity": "LOW",
|
|
"privilegesRequired": "NONE",
|
|
"userInteraction": "NONE",
|
|
"scope": "UNCHANGED",
|
|
"confidentialityImpact": "HIGH",
|
|
"integrityImpact": "NONE",
|
|
"availabilityImpact": "NONE",
|
|
"baseScore": 6.5,
|
|
"baseSeverity": "MEDIUM"
|
|
},
|
|
"exploitabilityScore": 2.8,
|
|
"impactScore": 3.6
|
|
}
|
|
],
|
|
"cvssMetricV2": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"cvssData": {
|
|
"version": "2.0",
|
|
"vectorString": "AV:A/AC:L/Au:N/C:P/I:N/A:N",
|
|
"accessVector": "ADJACENT_NETWORK",
|
|
"accessComplexity": "LOW",
|
|
"authentication": "NONE",
|
|
"confidentialityImpact": "PARTIAL",
|
|
"integrityImpact": "NONE",
|
|
"availabilityImpact": "NONE",
|
|
"baseScore": 3.3
|
|
},
|
|
"baseSeverity": "LOW",
|
|
"exploitabilityScore": 6.5,
|
|
"impactScore": 2.9,
|
|
"acInsufInfo": false,
|
|
"obtainAllPrivilege": false,
|
|
"obtainUserPrivilege": false,
|
|
"obtainOtherPrivilege": false,
|
|
"userInteractionRequired": false
|
|
}
|
|
]
|
|
},
|
|
"weaknesses": [
|
|
{
|
|
"source": "nvd@nist.gov",
|
|
"type": "Primary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "NVD-CWE-noinfo"
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"configurations": [
|
|
{
|
|
"operator": "AND",
|
|
"nodes": [
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": true,
|
|
"criteria": "cpe:2.3:o:drtrust:electrocardiogram_pen_firmware:2.00.08:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "DEC90897-AD8D-4072-A28E-6326302F64EE"
|
|
}
|
|
]
|
|
},
|
|
{
|
|
"operator": "OR",
|
|
"negate": false,
|
|
"cpeMatch": [
|
|
{
|
|
"vulnerable": false,
|
|
"criteria": "cpe:2.3:h:drtrust:electrocardiogram_pen:2.00.08:*:*:*:*:*:*:*",
|
|
"matchCriteriaId": "33AFB885-737E-419E-AB0B-905B7B54D700"
|
|
}
|
|
]
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"references": [
|
|
{
|
|
"url": "https://payatu.com/advisory/lack-of-bluetooth-le-encryption-and-access-control-in-dr-trust-ecg-or-ekg-pen",
|
|
"source": "cve@mitre.org",
|
|
"tags": [
|
|
"Third Party Advisory"
|
|
]
|
|
}
|
|
]
|
|
} |