"id": "CVE-2024-48967",
|
|
"sourceIdentifier": "productsecurity@baxter.com",
|
|
"published": "2024-11-14T22:15:17.927",
|
|
"lastModified": "2024-11-15T13:58:08.913",
|
|
"vulnStatus": "Awaiting Analysis",
|
|
"cveTags": [],
|
|
"descriptions": [
|
|
{
|
|
"lang": "en",
|
|
"value": "The ventilator and the Service PC lack sufficient audit logging capabilities to allow for detection of malicious activity and subsequent forensic examination. An attacker with access to the ventilator and/or the Service PC could, without detection, make unauthorized changes to ventilator settings that result in unauthorized disclosure of information and/or have unintended impacts on device performance."
|
|
},
|
|
{
|
|
"lang": "es",
|
|
"value": "El respirador y la PC de servicio carecen de capacidades de registro de auditor\u00eda suficientes para permitir la detecci\u00f3n de actividad maliciosa y el posterior examen forense. Un atacante con acceso al respirador o a la PC de servicio podr\u00eda, sin ser detectado, realizar cambios no autorizados en la configuraci\u00f3n del respirador que resulten en la divulgaci\u00f3n no autorizada de informaci\u00f3n o tengan impactos no deseados en el rendimiento del dispositivo."
|
|
}
|
|
],
|
|
"metrics": {
|
|
"cvssMetricV31": [
|
|
{
|
|
"source": "productsecurity@baxter.com",
|
|
"type": "Secondary",
|
|
"cvssData": {
|
|
"version": "3.1",
|
|
"vectorString": "CVSS:3.1/AV:N/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
|
|
"baseScore": 10.0,
|
|
"baseSeverity": "CRITICAL",
|
|
"attackVector": "NETWORK",
|
|
"attackComplexity": "LOW",
|
|
"privilegesRequired": "NONE",
|
|
"userInteraction": "NONE",
|
|
"scope": "CHANGED",
|
|
"confidentialityImpact": "HIGH",
|
|
"integrityImpact": "HIGH",
|
|
"availabilityImpact": "HIGH"
|
|
},
|
|
"exploitabilityScore": 3.9,
|
|
"impactScore": 6.0
|
|
}
|
|
]
|
|
},
|
|
"weaknesses": [
|
|
{
|
|
"source": "productsecurity@baxter.com",
|
|
"type": "Secondary",
|
|
"description": [
|
|
{
|
|
"lang": "en",
|
|
"value": "CWE-778"
|
|
}
|
|
]
|
|
}
|
|
],
|
|
"references": [
|
|
{
|
|
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01",
|
|
"source": "productsecurity@baxter.com"
|
|
}
|
|
]
|
|
} |