"id": "CVE-2024-9832",
"sourceIdentifier": "productsecurity@baxter.com",
"published": "2024-11-14T21:15:22.593",
"lastModified": "2024-11-15T13:58:08.913",
"vulnStatus": "Awaiting Analysis",
"cveTags": [],
"descriptions": [
{
"lang": "en",
"value": "There is no limit on the number of failed login attempts permitted with the Clinician Password or the Serial Number Clinician Password. An attacker could execute a brute-force attack to gain unauthorized access to the ventilator, and then make changes to device settings that could disrupt the function of the device and/or result in unauthorized information disclosure."
},
{
"lang": "es",
"value": "No existe l\u00edmite en la cantidad de intentos fallidos de inicio de sesi\u00f3n permitidos con la contrase\u00f1a del m\u00e9dico o la contrase\u00f1a del m\u00e9dico con n\u00famero de serie. Un atacante podr\u00eda ejecutar un ataque de fuerza bruta para obtener acceso no autorizado al respirador y luego realizar cambios en la configuraci\u00f3n del dispositivo que podr\u00edan interrumpir el funcionamiento del dispositivo o dar como resultado la divulgaci\u00f3n no autorizada de informaci\u00f3n."
}
],
"metrics": {
"cvssMetricV31": [
{
"source": "productsecurity@baxter.com",
"type": "Secondary",
"cvssData": {
"version": "3.1",
"vectorString": "CVSS:3.1/AV:L/AC:L/PR:N/UI:N/S:C/C:H/I:H/A:H",
"baseScore": 9.3,
"baseSeverity": "CRITICAL",
"attackVector": "LOCAL",
"attackComplexity": "LOW",
"privilegesRequired": "NONE",
"userInteraction": "NONE",
"scope": "CHANGED",
"confidentialityImpact": "HIGH",
"integrityImpact": "HIGH",
"availabilityImpact": "HIGH"
},
"exploitabilityScore": 2.5,
"impactScore": 6.0
}
]
},
"weaknesses": [
{
"source": "productsecurity@baxter.com",
"type": "Secondary",
"description": [
{
"lang": "en",
"value": "CWE-307"
}
]
}
],
"references": [
{
"url": "https://www.cisa.gov/news-events/ics-medical-advisories/icsma-24-319-01",
"source": "productsecurity@baxter.com"
}
]
}